Dosage-form testing
Liquid Particle Counter (Light Obscuration)
Counts and sizes sub-visible particles in injections by passing the liquid through a narrow light beam. Each particle casts a shadow whose size gives the particle size. Used for the USP <788> particulate matter test.
- Pharmaceutical (parenterals)
- Biotech
- Medical devices
- Industrial fluids
TROUBLESHOOTING
Common problems, causes and fixes
High blank counts
Likely causes
- • Dirty glassware or water
- • Contaminated sensor
- • Poor environment
What to do
- • Re-clean glassware; use freshly filtered water
- • Flush the sensor with detergent then water
- • Work in LAF and wear gloves
Counts falling between portions
Likely causes
- • Bubbles settling or dissolving
- • Particles settling
What to do
- • Degas properly before running
- • Stir gently during measurement
Very high counts at small sizes
Likely causes
- • Air bubbles
- • Product precipitating
What to do
- • Degas and re-run
- • Check product compatibility; consider the microscopic method
Count accuracy check fails
Likely causes
- • Coincidence errors
- • Wrong volume
- • Sensor drift
What to do
- • Check dilution
- • Verify syringe volume
- • Recalibrate the sensor