Dosage-form testing
Liquid Particle Counter (Light Obscuration)
Counts and sizes sub-visible particles in injections by passing the liquid through a narrow light beam. Each particle casts a shadow whose size gives the particle size. Used for the USP <788> particulate matter test.
- Pharmaceutical (parenterals)
- Biotech
- Medical devices
- Industrial fluids
Follow your site SOP
This is a typical operating sequence for learning. Your instrument model and approved SOP take precedence.
STAGE 1
Environment and blank
- 1Work in a laminar air flow cabinet; wear powder-free gloves.
- 2Clean glassware with particle-free water (filtered through 1.2 µm or finer).
- 3Run a blank of particle-free water: it should meet the environment limit in the chapter (commonly no more than 25 particles ≥ 10 µm in 25 mL).
- Most failures come from contamination by the analyst or glassware, not the product.
STAGE 2
Sample preparation
- 1Wipe the outside of containers and invert them gently about 20 times to mix.
- 2Open carefully; pool small units if the volume per container is too small for testing.
- 3Allow the pooled sample to stand or apply gentle vacuum so bubbles disappear.
- Do not shake hard; it makes bubbles that are counted as particles.
STAGE 3
Measurement
- 1Run four portions of at least 5 mL each.
- 2Discard the result of the first portion.
- 3Average the counts of the remaining portions.
- 4Calculate particles per container (small volume) or per mL (large volume) and compare with the limits.
- If counts are near the coincidence limit, the result is not valid; use a different volume or the microscopic method.
In LimsPulse · Calibration Management
Keep every Liquid particle counter in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.