LimsPulse
Dosage-form testing

Tablet Hardness Tester (Breaking Force Tester)

Measures the force needed to break a tablet by squeezing it between two jaws. Used to control compression and to make sure tablets are strong enough to survive coating, packing and handling.

  • Pharmaceutical
  • Nutraceuticals
  • Confectionery and food tablets
  • Veterinary medicines
Illustration of Tablet Hardness Tester (Breaking Force Tester)
Follow your site SOP
This is a typical operating sequence for learning. Your instrument model and approved SOP take precedence.
STAGE 1

Before you start

  1. 1Check the calibration status label.
  2. 2Clean the jaws so no powder or fragments remain.
  3. 3Select the correct units (N) and test settings for the product.
  • Never test with powder on the jaws; it changes contact and gives wrong values.
STAGE 2

Testing

  1. 1Take the number of tablets stated in the procedure (often 10).
  2. 2Place each tablet in the same orientation, centred between the jaws.
  3. 3Start the test and record the breaking force.
  4. 4Remove fragments and clean the jaws between tablets.
  5. 5Report individual values, mean, minimum and maximum.
  • Keep fingers clear of the jaws during the stroke.
STAGE 3

After testing

  1. 1Clean the jaws and the fragment tray
  2. 2Complete the usage log
In LimsPulse · Calibration Management

Keep every Hardness tester in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.

See how