LimsPulse
UV-Vis · Calibration procedure

Photometric linearity

Checks that absorbance rises in a straight line with concentration across the working range, using a series of potassium dichromate solutions.

Photo of a UV-Vis
Photo: Wiki210397, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Most UV assays compare a sample with a single standard or use a specific absorbance value. Both assume the Beer–Lambert law holds over the range measured.

Frequency

Typically every 6 months (some labs quarterly), and after lamp replacement, major repair or relocation, as defined in your calibration SOP.

Learn how the UV-Vis works
YOU WILL NEED

Standards & tools

  • Potassium dichromate solutions in 0.005 M sulfuric acid, at least five concentrations (e.g. 20, 40, 60, 80 and 100 mg/L)
  • Or a set of certified neutral-density filters covering about 0.2 to 2 A
BEFORE YOU START

Prerequisites

  • Wavelength and absorbance accuracy already passed
  • Solutions prepared by accurate dilution of one stock with Class A glassware
PROCEDURE

Step by step

  1. 1Zero the instrument with the acid blank at 350 nm (or the wavelength in your SOP).
  2. 2Measure each solution from the lowest to the highest concentration, rinsing the cuvette with the next solution each time.
  3. 3Plot absorbance against concentration and calculate the regression line and the correlation coefficient.
  4. 4Optionally check that the specific absorbance is constant at each level.
RECORD

Example observation table

Concentration (mg/L)Absorbance at 350 nmA(1%, 1 cm)
200.214107.0
400.427106.8
600.640106.7
800.851106.4
1001.067106.7
CALCULATION

How to calculate the result

Linear regression A = m × c + intercept; report the correlation coefficient r
  1. Using the five points above: slope m = 0.01065 A per mg/L, intercept = 0.0008
  2. r = 0.99999
  3. Specific absorbance stays between 106.4 and 107.0 at every level, confirming a straight-line response
Use the Standard curve (linear regression) calculator
ACCEPTANCE CRITERIA

Pass if

  • Correlation coefficient r typically not less than 0.999 over the tested range.
  • Where filters with certified values are used, each reading within its certified tolerance (as per your SOP).
IF IT FAILS

What to do

  • Check solution preparation and cuvette cleanliness; repeat once if an error is documented.
  • Check stray light, since high stray light causes curvature at the top of the range.
  • Limit the working range in use or arrange service if it still fails.
MORE

Other UV-Vis calibrations

In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how