UV-Vis · Calibration procedure

Absorbance accuracy

Checks that the absorbance reading is correct, by measuring a potassium dichromate solution of known concentration in dilute acid and comparing the specific absorbance with reference values.

Photo of a UV-Vis
Photo: Wiki210397, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Assays calculated from a specific absorbance A(1%, 1 cm), and any test with an absorbance limit, depend directly on the photometric scale being right. Absorbance accuracy is sometimes called photometric accuracy.

Frequency

Typically every 6 months (some labs quarterly), and after lamp replacement, major repair or relocation, as defined in your calibration SOP.

Learn how the UV-Vis works
YOU WILL NEED

Standards & tools

  • Potassium dichromate (CRM or primary-standard grade), dried at about 130 °C to constant weight
  • 0.005 M sulfuric acid (as in Ph. Eur.) or 0.001 M perchloric acid (as in some USP procedures) as solvent and blank
  • Alternatively, certified neutral-density glass filters or sealed liquid CRMs with certified absorbance values
  • Matched 10 mm quartz cuvettes
BEFORE YOU START

Prerequisites

  • Wavelength accuracy already passed
  • Calibrated analytical balance and Class A volumetric glassware
  • Lamps warmed up; cuvettes clean and matched
PROCEDURE

Step by step

  1. 1Accurately weigh 57.0–63.0 mg of dried potassium dichromate and dissolve it in the acid to make 1000 mL (about 60 mg/L).
  2. 2Zero the instrument with the acid blank at each test wavelength.
  3. 3Measure the absorbance of the solution at 235, 257, 313 and 350 nm (the maxima and minima of the dichromate spectrum).
  4. 4Convert each absorbance into specific absorbance using the exact concentration.
  5. 5Compare with the reference values and tolerances given in the current chapter or on the CRM certificate.
  6. 6If your SOP includes 430 nm, a more concentrated dichromate solution is used as the chapter describes.
RECORD

Example observation table

Wavelength (nm)AbsorbanceA(1%, 1 cm) foundTypical reference valueTypical tolerance
235 (min)0.748124.5124.5122.9 – 126.2
257 (max)0.868144.5144.5142.8 – 146.2
313 (min)0.29248.648.647.0 – 50.3
350 (max)0.640106.6106.6105.6 – 108.2
CALCULATION

How to calculate the result

A(1%, 1 cm) = A ÷ (c × b), where c is concentration in g/100 mL and b is path length in cm
  1. 60.06 mg of potassium dichromate in 1000 mL = 0.006006 g/100 mL
  2. Absorbance at 257 nm = 0.868, path length 1 cm
  3. A(1%, 1 cm) = 0.868 ÷ (0.006006 × 1) = 144.5
  4. 144.5 lies within 142.8 – 146.2, so 257 nm passes
ACCEPTANCE CRITERIA

Pass if

  • Specific absorbance at each wavelength within the tolerance in the current chapter (the typical values above), or within the certified value ± its stated tolerance for a CRM.
  • Where the chapter expresses the limit directly in absorbance, a typical tolerance is about ±0.010 A for readings up to 1.0 A (±1% above that); follow your SOP.
IF IT FAILS

What to do

  • Check weighing, drying and dilution records, cuvette cleanliness and the blank; repeat once if an error is found and documented.
  • Check stray light and lamp energy, which commonly cause low readings at low wavelengths.
  • Label the instrument 'Out of calibration' and arrange service if it still fails.
  • Assess the impact on assays reported since the last passing calibration.