Proves that high and low temperature and RH alarms, door alarms and power-failure alarms trigger at the right limits and reach the right people.

Photo of a Stability chamber
Photo: Encik Tekateki, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Alarms are the safety net for long stability studies. If an alarm fails, an excursion could last for days unnoticed and a whole study could be lost.

Frequency

Typically every 6 months and after any change to alarm settings, software or notification lists, as defined in your SOP.

Learn how the Stability chamber works
YOU WILL NEED

Standards & tools

  • Calibrated reference thermometer/hygrometer
  • Calibrated stopwatch
BEFORE YOU START

Prerequisites

  • Samples protected or moved if the test will push conditions outside limits
  • Alarm set points and delays documented in the SOP
PROCEDURE

Step by step

  1. 1Record the configured alarm limits (e.g. 38 °C and 42 °C, 70% RH and 80% RH at 40 °C/75% RH) and the alarm delay.
  2. 2Raise or lower the condition (or simulate with a calibrator, as the SOP allows) until the alarm limit is crossed.
  3. 3Note the reference reading when the alarm triggers and the time from crossing the limit to the alarm.
  4. 4Check that the local audible/visual alarm sounds and that remote notifications (SMS, e-mail, BMS) arrive.
  5. 5Test door-open and power-failure alarms the same way.
  6. 6Return the chamber to normal and confirm conditions recover.
RECORD

Example observation table

AlarmLimitReference at triggerDelay setDelay observedNotification
High temperature42.0 °C42.1 °C15 min15 min 20 sReceived
Low temperature38.0 °C37.9 °C15 min15 min 10 sReceived
High RH80% RH80.4% RH15 min15 min 30 sReceived
Door openOpen > 2 min—2 min2 min 05 sReceived
CALCULATION

How to calculate the result

Trigger error = Reference reading at alarm − Alarm limit
  1. High temperature limit 42.0 °C; the alarm triggered when the reference read 42.1 °C
  2. Trigger error = 42.1 − 42.0 = +0.1 °C, within the sensor tolerance
  3. Delay observed 15 min 20 s against 15 min set: alarm works as configured
ACCEPTANCE CRITERIA

Pass if

  • Every alarm triggers at its limit within the sensor tolerance and after the configured delay.
  • Local and remote notifications reach all people on the call-out list.
IF IT FAILS

What to do

  • Treat the chamber as unmonitored: increase manual checks or move samples.
  • Correct the alarm configuration or hardware and repeat the test.
  • Raise a deviation and assess whether any excursions may have gone unnoticed.