LimsPulse
FTIR · Calibration procedure

Wavenumber accuracy (polystyrene)

Checks that the FTIR reports absorption bands at the correct wavenumbers, by recording the spectrum of a certified polystyrene film and comparing band positions with reference values.

Photo of a FTIR
Photo: Wang, F., Zhang, L., Zhang, X., Li, H. and Wu, S., CC BY 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Identification compares the positions of bands in sample and reference spectra. A shifted wavenumber scale could make a correct material fail, or hide a real difference.

Frequency

Full calibration typically every 6 months or yearly; a quick polystyrene check is often done daily or before use, as defined in your SOP.

Learn how the FTIR works
YOU WILL NEED

Standards & tools

  • Polystyrene film, about 35–50 µm thick, with a certificate of band positions traceable to a national standard
BEFORE YOU START

Prerequisites

  • Instrument warmed up and humidity indicator acceptable
  • Resolution and apodisation set as stated in your SOP or the film certificate (commonly 2 or 4 cm⁻¹)
  • Sample compartment closed long enough for the atmosphere to settle
PROCEDURE

Step by step

  1. 1Collect a background spectrum with the beam path empty (transmission holder).
  2. 2Place the polystyrene film in the holder and record its spectrum over about 4000–400 cm⁻¹.
  3. 3Use the software's peak-pick to locate the band minima in %T (maxima in absorbance) at the certified positions.
  4. 4Compare each observed position with the certified value and calculate the difference.
  5. 5Do not use an ATR spectrum for this test unless your SOP and the certificate cover ATR, because ATR shifts some band positions slightly.
RECORD

Example observation table

Typical reference band (cm⁻¹)Observed (cm⁻¹)Difference (cm⁻¹)Typical tolerance (cm⁻¹)
3060.03060.3+0.3±1.5
2849.52849.9+0.4±2.0
1942.91942.6−0.3±1.5
1601.21601.4+0.2±1.0
1583.01583.1+0.1±1.0
1154.51154.4−0.1±1.0
1028.31028.1−0.2±1.0
CALCULATION

How to calculate the result

Difference = Observed wavenumber − Reference wavenumber
  1. Reference band 1601.2 cm⁻¹, observed 1601.4 cm⁻¹
  2. Difference = 1601.4 − 1601.2 = +0.2 cm⁻¹
  3. Within the typical ±1.0 cm⁻¹ tolerance, so the band passes
ACCEPTANCE CRITERIA

Pass if

  • Each band within the tolerance in the current Ph. Eur. 2.2.24 / USP <854> table or the film certificate (typically ±1.0 cm⁻¹ for sharp bands in the fingerprint region, ±1.5 to ±2.0 cm⁻¹ for bands above 1900 cm⁻¹).
IF IT FAILS

What to do

  • Check the film for damage and the method settings (resolution, apodisation); repeat once if an error is documented.
  • Stop using the instrument and label it 'Out of calibration'.
  • Arrange service to check the reference laser and alignment.
  • Assess identification results since the last passing check.
MORE

Other FTIR calibrations

In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how