FTIR · Calibration procedure

Signal-to-noise

Measures the noise on a '100% line' (two backgrounds ratioed against each other) and converts it into a signal-to-noise ratio to compare with the manufacturer's specification.

Photo of a FTIR
Photo: Wang, F., Zhang, L., Zhang, X., Li, H. and Wu, S., CC BY 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Low signal-to-noise hides weak bands and makes library matching less reliable. A falling ratio is often the first sign of a weak source, a fogged beam splitter or a failing detector.

Frequency

Full calibration typically every 6 months or yearly; a quick polystyrene check is often done daily or before use, as defined in your SOP.

Learn how the FTIR works
YOU WILL NEED

Standards & tools

  • No physical standard: the test uses the empty beam (or a clean ATR crystal) under defined conditions
  • Practice described in ASTM E1421 and the manufacturer's OQ procedure
BEFORE YOU START

Prerequisites

  • Instrument fully warmed up and stable
  • Conditions fixed as in the specification (e.g. resolution 4 cm⁻¹, a stated scan time or number of scans, a stated wavenumber region)
PROCEDURE

Step by step

  1. 1Collect a background spectrum with the empty beam.
  2. 2Immediately collect a second spectrum under the same conditions and ratio it against the first to give a 100% transmittance line.
  3. 3Measure the peak-to-peak (or RMS) noise in %T over the specified region (e.g. 2200–2000 cm⁻¹, where there are no atmospheric bands).
  4. 4Calculate signal-to-noise and repeat as many times as your SOP requires.
RECORD

Example observation table

RunRegion (cm⁻¹)Peak-to-peak noise (%T)S/N (peak-to-peak)
12200–20000.003033,333 : 1
22200–20000.002835,714 : 1
CALCULATION

How to calculate the result

S/N = 100 ÷ peak-to-peak noise (in %T) on the 100% line
  1. Peak-to-peak noise over 2200–2000 cm⁻¹ = 0.0030 %T
  2. S/N = 100 ÷ 0.0030 = 33,333 : 1
  3. Compare with the manufacturer's specification for the same resolution, scan time and region
ACCEPTANCE CRITERIA

Pass if

  • Signal-to-noise not less than the manufacturer's specification (or your SOP limit) for the stated conditions. Values differ widely between models and conditions, so there is no single compendial number.
IF IT FAILS

What to do

  • Check the desiccant and humidity; replace if exhausted.
  • Check source energy and interferogram amplitude; run alignment.
  • Arrange service for source, beam splitter or detector if it still fails.