LimsPulse
Ambient air · Particulate matter collected on filters over 24 hours

PM10 and PM2.5 in Ambient Air by Gravimetric Sampling

Measures the mass concentration of airborne particles in two size fractions. Air is drawn at a known, controlled flow through a size-selective inlet for 24 hours; the particles that pass the inlet collect on a pre-weighed filter. The filter is conditioned and re-weighed, and the mass gain divided by the volume of air sampled gives µg/m³.

  • Environmental & water
  • Size-selective inlet, controlled-flow sampling onto a filter, and weighing of conditioned filters
  • PM sampler
  • Microbalance
  • Analytical balance

Reference: Typical procedure based on CPCB Guidelines for the Measurement of Ambient Air Pollutants (Vol. I), IS 5182 (Part 23) for PM10, and US EPA 40 CFR Part 50 Appendix J (PM10) and Appendix L (PM2.5). Follow the method approved for your station.

PRINCIPLE

Why this method works

An impactor or cyclone inlet removes particles larger than the cut point (10 µm or 2.5 µm aerodynamic diameter) because their inertia makes them strike a surface instead of following the air stream. The remaining particles are captured on a filter: glass-fibre or quartz filters for high-volume PM10 samplers, and 47 mm PTFE (Teflon) membrane filters for low-volume PM2.5 samplers. Filters are weighed before and after sampling after equilibrating in a room with controlled temperature and humidity, because filters and particles gain or lose water with humidity. Concentration = mass gain ÷ volume of air sampled. The flow rate must be held at the design value because the size cut of the inlet depends on it.

Instruments used

MATERIALS

Reagents & standards

ReagentGrade / specification
PM2.5 filters47 mm PTFE membrane filters with support ring, low background, numbered
PM10 filtersGlass-fibre (for example 20 × 25 cm for high-volume samplers) or PTFE/quartz for low-volume samplers
Check weightsCalibrated working mass standards (for example 100 mg and 200 mg) for the balance
Inlet grease / oilImpactor oil or grease as specified by the sampler maker (for example for a WINS impactor)
ConsumablesAntistatic device (ionizer or polonium strip), non-serrated forceps, labelled filter cassettes and petri slides
SAMPLE PREPARATION

Preparing the sample

  1. 01

    Inspect filters

    Check each filter for pinholes, tears or loose material.

  2. 02

    Condition

    Equilibrate at controlled temperature and RH, typically at least 24 h.

  3. 03

    Pre-weigh

    Weigh with static removed; record initial mass.

  4. 04

    Sample

    Run the sampler at design flow for 24 h; record flow, time, T and P.

  5. 05

    Transport

    Return the filter in its cassette, protected and cool.

  6. 06

    Post-weigh

    Re-condition and re-weigh; mass gain ÷ air volume.

Procedure

  1. 1Inspect each filter against a light. Reject filters with pinholes, tears, chaff or flashing. Number each filter and its holder.
  2. 2Condition the filters in the weighing room. Typical conditions are a mean temperature of about 20–23 °C and a mean relative humidity of about 30–40% (PM2.5), held stable over at least 24 hours. Use the conditions in your method.
  3. 3Zero and check the balance with the working mass standards. Remove static from each filter and weigh it (microbalance with 1 µg readability for 47 mm PM2.5 filters; analytical balance with 0.1 mg readability for large PM10 glass-fibre filters). Weigh one or two laboratory blank filters with every batch.
  4. 4At the site, check the sampler flow against a calibrated flow transfer standard, clean and re-grease the impactor if needed, and install the filter cassette without touching the filter surface.
  5. 5Run the sampler for 24 hours (typically from 00:00 to 00:00 or as per the monitoring schedule). Record the start and end times, elapsed time, average flow rate, ambient temperature and pressure, and any flow faults.
  6. 6Recover the filter, inspect it, place it in its labelled holder and transport it to the lab protected from heat and vibration. Carry a field blank filter that goes to the site but is not sampled.
  7. 7Re-condition the exposed filter under the same conditions as before and re-weigh it. Repeat weighings should agree within the tolerance in your SOP (for example within a few µg for PM2.5 filters).
  8. 8Calculate the volume of air sampled and the mass concentration. Archive the filter in a refrigerator if chemical analysis (metals, ions, carbon) may follow.

Standard preparation

  1. 1Weigh working mass standards at the start and end of each weighing session; they must read within the tolerance of their certified values.
  2. 2Weigh laboratory blank filters and field blanks with each batch; a large blank mass change shows a conditioning, handling or contamination problem.

Notes for analysts

  • Always state whether volumes are at actual or standard conditions; the method you follow defines which is used.
  • Static charge on PTFE filters causes drifting readings. Use an antistatic device and allow readings to stabilise.
  • Handle filters only with clean, non-serrated forceps, and wear powder-free gloves.
  • A blocked or under-greased impactor shifts the size cut. Clean the inlet on the schedule set by the maker.
  • Volatile components (ammonium nitrate, some organics) can be lost from filters in warm conditions; transport and store exposed filters cool.
TEST CONDITIONS

Method parameters (typical)

Sampling period24 hours (typically 1440 minutes ± a small allowance defined in the method)
PM2.5 low-volume flow16.67 L/min (1.0 m³/h) at actual conditions
PM10 high-volume (respirable dust sampler) flowTypically about 1.0–1.2 m³/min, with cyclone or impactor inlet
PM10 low-volume flow16.67 L/min with a PM10 inlet
Filter47 mm PTFE (PM2.5); glass-fibre or quartz (PM10 high-volume)
BalanceMicrobalance, 1 µg readability (47 mm filters); analytical balance, 0.1 mg (large filters)
ConditioningControlled temperature and RH, typically ≥ 24 h before each weighing
BEFORE YOU REPORT

System suitability & sequence

ParameterLimit
Flow checkSampler flow within about ±4% of the transfer standard and within about ±5% of the design flow (typical criteria; see your method)
Balance checkWorking standards within tolerance at start and end of session
Lab and field blanksMass change within the limit set in your SOP
Run validitySampling time and flow-stability criteria met; no unresolved sampler alarms
Collocated samplersPrecision between collocated samplers tracked over time

Typical sequence

  1. 1Balance zero and check weights
  2. 2Laboratory blank filters
  3. 3Pre-weigh sample filters
  4. 4Field flow check and sampling (24 h)
  5. 5Post-weigh exposed filters and field blanks
  6. 6Balance check weights at end of session
CALCULATION

Calculation with a worked example

Formula
C (µg/m³) = (W₂ − W₁) ÷ V; V (m³) = Q × t

Where

C
Mass concentration of PM (µg/m³)
W₁
Conditioned filter mass before sampling (µg)
W₂
Conditioned filter mass after sampling (µg)
Q
Average flow rate (m³/min)
t
Elapsed sampling time (min)
V
Volume of air sampled (m³), at the conditions required by your method

Given

PM2.5: W₁
148.235 mg (microbalance)
PM2.5: W₂
149.705 mg
PM2.5: Q, t
16.67 L/min for 1440 min
PM10 (high-volume): mass gain
0.1250 g on a glass-fibre filter
PM10: Q, t
1.10 m³/min for 1440 min

Working

  1. 1

    PM2.5 mass gain = 149.705 − 148.235 = 1.470 mg = 1470 µg

  2. 2

    PM2.5 air volume = 16.67 L/min × 1440 min = 24 004.8 L = 24.005 m³

  3. 3
    PM2.5 concentration = 1470 ÷ 24.005 = 61.2 µg/m³
  4. 4

    PM10 air volume = 1.10 × 1440 = 1584 m³

  5. 5
    PM10 concentration = 125 000 µg ÷ 1584 m³ = 78.9 µg/m³

PM2.5 = 61.2 µg/m³ (24-h); PM10 = 78.9 µg/m³ (24-h)

ACCEPTANCE CRITERIA

What result passes

  • Compare 24-hour results with the National Ambient Air Quality Standards (India). As notified in NAAQS 2009, the 24-hour limits are 100 µg/m³ for PM10 and 60 µg/m³ for PM2.5 (annual 60 and 40 µg/m³); check the current notification.
  • In the worked example, PM2.5 at 61.2 µg/m³ is above the 24-hour value of 60 µg/m³, while PM10 at 78.9 µg/m³ is below 100 µg/m³.
  • The notification's notes state that 24-hour values are to be complied with 98% of the time in a year, so compliance is judged over the monitoring record. A single result is reported, not 'passed' or 'failed'.
METHOD VALIDATION

Validation parameters (ICH Q2(R2))

ParameterTypical requirement
PrecisionCollocated sampler comparisons; replicate weighings of blanks
TruenessFlow-rate audits with an independent transfer standard; balance calibration with traceable weights
Detection limitFrom the variability of laboratory and field blank filters
ComparabilityAgainst a reference or equivalent method sampler where required
In LimsPulse · Sample Management

Run this method in LimsPulse: specifications and STPs under version control, results calculated and checked against limits, and a signed Certificate of Analysis.

See how