Physical testing
Turbidity meter
Measures how cloudy a liquid is by shining light through it and measuring the light scattered by tiny particles. Results are given in NTU or FNU and are used for clarity and opalescence tests.
- Pharmaceutical
- Biotech
- Water treatment
- Food & beverage
- Brewing
- Environmental testing
Follow your site SOP
This is a typical operating sequence for learning. Your instrument model and approved SOP take precedence.
STAGE 1
Before you start
- 1Check the calibration status and run the daily verification with a secondary standard.
- 2Switch on and let the lamp warm up as the manufacturer states.
- 3Clean the cuvettes inside and out; check for scratches.
- Handle cuvettes only by the top; fingerprints scatter light.
STAGE 2
Measurement
- 1Rinse the cuvette with the sample, then fill it to the mark.
- 2Cap it and wipe the outside with a lint-free cloth; apply a thin film of silicone oil if your SOP requires it to mask small scratches.
- 3Let air bubbles rise, or remove them by gentle inversion or a short time in an ultrasonic bath, as the method allows.
- 4Place the cuvette in the holder with the orientation mark aligned and close the lid.
- 5Record the stable reading in NTU or FNU.
- Do not shake the sample vigorously; it creates bubbles.
- Cold samples can fog the cuvette. Let them reach room temperature or dry the outside before reading.
STAGE 3
After use
- 1Empty and rinse the cuvette with purified water, then let it dry upside down.
- 2Store cuvettes and standards as the manufacturer directs.
- 3Complete the log.