Measures how cloudy a liquid is by shining light through it and measuring the light scattered by tiny particles. Results are given in NTU or FNU and are used for clarity and opalescence tests.

  • Pharmaceutical
  • Biotech
  • Water treatment
  • Food & beverage
  • Brewing
  • Environmental testing
Illustration of Turbidity meter
PREVENTIVE MAINTENANCE

Maintenance checklist

Daily / each use

  • Verify with a secondary standard
  • Clean cuvettes
  • Check the cuvette holder for dust

Monthly

  • Inspect cuvettes and replace scratched ones
  • Check expiry of standards
  • Clean the optical compartment

Quarterly / six-monthly

  • Calibrate with formazin standards (or as the manufacturer and your SOP require)
  • Re-assign values to secondary standards after calibration
  • Replace the lamp when its output is low
SAFETY

Safety precautions

Before you work on the Turbidity meter
  • Hydrazine sulphate, used to make formazin, is toxic and a suspected carcinogen. Prepare formazin only in a fume hood with gloves and eye protection, or use ready-made stabilised standards.
  • Dispose of formazin waste as hazardous chemical waste.
  • Handle glass cuvettes carefully to avoid breakage.