PREVENTIVE MAINTENANCE
Maintenance checklist
Daily / each use
- Verify with a secondary standard
- Clean cuvettes
- Check the cuvette holder for dust
Monthly
- Inspect cuvettes and replace scratched ones
- Check expiry of standards
- Clean the optical compartment
Quarterly / six-monthly
- Calibrate with formazin standards (or as the manufacturer and your SOP require)
- Re-assign values to secondary standards after calibration
- Replace the lamp when its output is low
SAFETY
Safety precautions
Before you work on the Turbidity meter
- Hydrazine sulphate, used to make formazin, is toxic and a suspected carcinogen. Prepare formazin only in a fume hood with gloves and eye protection, or use ready-made stabilised standards.
- Dispose of formazin waste as hazardous chemical waste.
- Handle glass cuvettes carefully to avoid breakage.