Titration & electrochemistry

Total Organic Carbon (TOC) Analyser

Oxidises organic compounds in water to carbon dioxide and measures the CO₂ formed to give the total organic carbon content. The standard test for organic contamination in purified water and water for injections.

  • Pharmaceutical
  • Biotech
  • Semiconductors
  • Power generation
  • Environmental
  • Drinking water
Follow your site SOP
This is a typical operating sequence for learning. Your instrument model and approved SOP take precedence.
STAGE 1

Before you start

  1. 1Check the analyser's calibration and last system suitability are within date.
  2. 2Check reagent levels (acid, oxidiser) and gas supply if used.
  3. 3Use TOC-certified (low-carbon) vials and caps; do not touch the inside of vials or caps.
  • Wear powder-free gloves and avoid solvents or perfumes near the samples: they add organic carbon.
STAGE 2

System suitability

  1. 1Analyse the reagent water blank, the 0.50 mg/L carbon sucrose standard and the 0.50 mg/L carbon 1,4-benzoquinone solution.
  2. 2Calculate the response efficiency and check it is within 85–115%.
STAGE 3

Sample analysis

  1. 1Collect samples in clean TOC vials, filling them with no headspace if required by the SOP.
  2. 2Analyse promptly, or store as validated (for example refrigerated for a defined time).
  3. 3Run samples, with a blank and a check standard as required by the sequence.
  4. 4Compare each sample result with the limit response.
  • Avoid air bubbles and headspace: CO₂ and organic vapours from air can change the result.
STAGE 4

Shutdown

  1. 1Flush the system with reagent water as recommended by the supplier.
  2. 2Put the instrument in standby and complete the usage log.
In LimsPulse · Calibration Management

Keep every TOC analyser in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.