LimsPulse
Titration & electrochemistry

Total Organic Carbon (TOC) Analyser

Oxidises organic compounds in water to carbon dioxide and measures the CO₂ formed to give the total organic carbon content. The standard test for organic contamination in purified water and water for injections.

  • Pharmaceutical
  • Biotech
  • Semiconductors
  • Power generation
  • Environmental
  • Drinking water
CALIBRATION

TOC analyser calibration procedures

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    DQ — Design qualification

    Defines the requirements: offline or online use, range (for example ppb levels), oxidation and detection type, compliance with USP <643> and data integrity features.

  2. 02

    IQ — Installation qualification

    Confirms installation, utilities (power, gas, sample lines), serial numbers, software version and documentation.

  3. 03

    OQ — Operational qualification

    Verifies calibration accuracy, linearity, repeatability, system suitability with sucrose and benzoquinone, and limit of detection.

  4. 04

    PQ — Performance qualification

    Shows routine performance over time, for example passing system suitability and stable blanks; for online units, comparison with offline grab samples.

In LimsPulse · Calibration Management

Keep every TOC analyser in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.

See how