Thermal analysis & ovens

Muffle Furnace

An electrically heated, insulated chamber that reaches several hundred to over 1000 °C. In QC it is used to ignite samples for sulphated ash (residue on ignition) and to bring crucibles to constant mass.

  • Pharmaceutical
  • Food
  • Cement and ceramics
  • Mining
  • Environmental
Photo of a Muffle Furnace
Photo: Karelj, Public domain (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

Slow to reach set point
Likely causes
  • • Ageing heating element
  • • Door seal damaged
  • • Supply voltage low
What to do
  • • Have elements checked and replaced
  • • Replace the seal
  • • Check the electrical supply
Temperature overshoots
Likely causes
  • • Controller tuning poor
  • • Thermocouple loose or moved
What to do
  • • Re-tune the PID controller
  • • Refit the thermocouple
Display differs from reference thermocouple
Likely causes
  • • Controller thermocouple drifted
  • • Probe placed in a cooler zone
What to do
  • • Calibrate or replace the thermocouple
  • • Measure in the working zone
Sulphated ash results variable
Likely causes
  • • Crucibles placed near the door
  • • Cooling time not controlled
  • • Incomplete charring
What to do
  • • Use the central working zone
  • • Use a fixed desiccator cooling time
  • • Char completely before ignition