Vacuum oven · Calibration procedure

Temperature accuracy

Compares the oven display with a calibrated temperature probe on the shelf at the drying temperatures used.

Photo of a Vacuum oven
Photo: GOKLuLe 盧樂, CC BY-SA 3.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

LOD in vacuum specifies a drying temperature. The shelf where samples sit must be at that temperature.

Frequency

Typically every 6 months, as defined in your SOP.

Learn how the Vacuum oven works
YOU WILL NEED

Standards & tools

  • Calibrated digital thermometer with a surface or air probe, traceable to national standards
BEFORE YOU START

Prerequisites

  • Probe lead passed through a vacuum-tight port if testing under vacuum
  • Oven stabilised at each set point
PROCEDURE

Step by step

  1. 1Place the probe on the middle shelf.
  2. 2Set the first temperature (for example 40 °C) and allow it to stabilise for at least 30 minutes.
  3. 3Record display and reference readings.
  4. 4Repeat at the other set points (for example 60 °C and 80 °C).
RECORD

Example observation table

Set point (°C)Reference (°C)Difference (°C)
4039.4−0.6
6060.8+0.8
8079.1−0.9
CALCULATION

How to calculate the result

Difference = Reference reading − Set point
  1. At 60 °C: 60.8 − 60 = +0.8 °C
ACCEPTANCE CRITERIA

Pass if

  • Difference within ±2 °C (typical; check your SOP).
IF IT FAILS

What to do

  • Label the instrument 'Out of calibration' and stop using it.
  • Investigate the cause and arrange adjustment or repair.
  • Assess the impact on results generated since the last passing calibration.
  • Repeat the full calibration before returning it to use.