Sieve shaker · Calibration procedure

Sieve aperture verification

Checks that the mesh openings of each test sieve are within the tolerances for its nominal size, by optical measurement or with certified calibration material.

Photo of a Sieve shaker
Photo: Cjp24, CC BY-SA 3.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Sieve meshes stretch, dent and wear with use. A sieve with enlarged openings lets coarse particles through and makes the product look finer than it is.

Frequency

Typically annually (more often for heavily used fine sieves), and whenever damage is suspected, as defined in your SOP.

Learn how the Sieve shaker works
YOU WILL NEED

Standards & tools

  • Calibrated measuring microscope or optical sieve analyser with a traceable scale
  • Or certified calibration spheres / reference material for the sieve size
  • Supplier's certificate of compliance with ISO 3310-1 or ASTM E11
BEFORE YOU START

Prerequisites

  • Sieve clean and dry
  • Visual inspection done: no holes, tears, loose wires or frame damage
PROCEDURE

Step by step

  1. 1Inspect the mesh against a light for damage; reject damaged sieves straight away.
  2. 2Measure a set number of apertures in both directions at several locations across the mesh (as per the standard or SOP).
  3. 3Calculate the mean aperture and note the largest single aperture.
  4. 4Compare the mean with the nominal size and the tolerances in ISO 3310-1 / ASTM E11.
  5. 5Record the result against the sieve serial number.
RECORD

Example observation table

Sieve (nominal)Apertures measuredMean (µm)Largest (µm)Deviation of mean
250 µm248, 255, 251, 253, 250, 256, 249, 252, 254, 251251.9256+1.9 µm (+0.76%)
CALCULATION

How to calculate the result

Mean aperture = Σ measured apertures ÷ n; Deviation = Mean − Nominal
  1. Sum of 10 apertures = 2519 µm
  2. Mean = 2519 ÷ 10 = 251.9 µm
  3. Deviation = 251.9 − 250 = +1.9 µm (+0.76%)
ACCEPTANCE CRITERIA

Pass if

  • Mean aperture and largest individual aperture within the ISO 3310-1 / ASTM E11 tolerances for that nominal size (see the standard's table).
  • No visible damage, holes or loose wires.
IF IT FAILS

What to do

  • Withdraw the sieve from use and label it rejected.
  • Replace with a certified sieve.
  • Assess results obtained with that sieve since the last verification.