LimsPulse
Potentiometric titrator · Calibration procedure

Electrode check

Checks a pH glass electrode's slope and zero point with standard buffers before it is used for acid–base titrations.

Photo of a Potentiometric titrator
Photo: User:FA2010, Public domain (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

An ageing or dirty electrode responds slowly and with a reduced slope, which flattens the titration curve and makes the equivalence point less sharp. The check confirms the electrode responds as expected.

Frequency

Typically daily or before use for pH-based titrations, as defined in your SOP.

Learn how the Potentiometric titrator works
YOU WILL NEED

Standards & tools

  • Certified pH buffers traceable to national standards, for example pH 4.01, 7.00 and 9.21 at 25 °C
  • Calibrated temperature sensor
BEFORE YOU START

Prerequisites

  • Electrolyte filled to the correct level and refill opening open
  • Buffers at about 25 °C or temperature compensation active
  • Electrode rinsed with water between buffers
PROCEDURE

Step by step

  1. 1Immerse the electrode in pH 7.00 buffer with gentle stirring and record the stable potential (mV).
  2. 2Rinse, then repeat with pH 4.01 and pH 9.21 buffers.
  3. 3Note the time each reading takes to stabilise (response time).
  4. 4Calculate the slope and express it as a percentage of the theoretical Nernst slope.
RECORD

Example observation table

Buffer (pH at 25 °C)Potential (mV)Response time (s)
4.01+172.518
7.00+0.615
9.21−126.422
CALCULATION

How to calculate the result

Slope (mV/pH) = (E₁ − E₂) ÷ (pH₂ − pH₁); Slope % = Slope ÷ 59.16 × 100 (at 25 °C)
  1. Slope = (172.5 − (−126.4)) ÷ (9.21 − 4.01) = 298.9 ÷ 5.20 = 57.48 mV/pH
  2. Slope % = 57.48 ÷ 59.16 × 100 = 97.2%
  3. Zero point: +0.6 mV at pH 7.00
ACCEPTANCE CRITERIA

Pass if

  • Slope typically 95–105% of the theoretical value.
  • Potential at pH 7.00 typically within ±15 mV (use the limit in your SOP).
  • Stable reading within about 30–60 seconds in each buffer.
IF IT FAILS

What to do

  • Refill or replace the reference electrolyte and clean the diaphragm.
  • Clean or recondition the glass membrane as per the supplier's instructions.
  • Repeat with fresh buffers; replace the electrode if it still fails.
MORE

Other Potentiometric titrator calibrations

In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how