Potentiometric titrator · Calibration procedure

Burette accuracy

Checks the delivered volume of the titrator's burette gravimetrically with water, corrected for temperature using the Z factor.

Photo of a Potentiometric titrator
Photo: User:FA2010, Public domain (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

The titrant volume at the equivalence point drives every assay result. A burette that delivers too much or too little gives a proportional error in the result.

Frequency

Typically every 6 or 12 months and after repair, as defined in your calibration SOP.

Learn how the Potentiometric titrator works
YOU WILL NEED

Standards & tools

  • Calibrated analytical balance (readability 0.1 mg or better)
  • Degassed purified water
  • Calibrated thermometer
  • Z factor table (mL/g) for water at the test temperature
BEFORE YOU START

Prerequisites

  • Burette unit flushed and filled with water
  • Water, balance and burette at room temperature
PROCEDURE

Step by step

  1. 1Fill the burette with degassed purified water and run several prepare / flush cycles to remove air bubbles.
  2. 2Record the water temperature and note the matching Z factor (mL/g) from the table in your SOP (about 1.0033 mL/g at 22 °C).
  3. 3Place a weighing vessel with a little water (to reduce evaporation) on the balance and tare it.
  4. 4Dispense the first test volume from the burette into the vessel, touching the tip to the wall to remove the last drop, and record the mass.
  5. 5Repeat for each test volume and replicate required by the SOP (for example 10%, 50% and 100% of the burette volume).
  6. 6Convert each mass to volume with the Z factor and compare it with the nominal volume.
RECORD

Example observation table

Nominal (mL)Mass (g)Z (mL/g, 22 °C)Volume (mL)Error (mL)Error (%)
5.0004.98611.00335.0026+0.0026+0.05
10.0009.96881.003310.0017+0.0017+0.02
15.00014.95471.003315.0041+0.0041+0.03
20.00019.94021.003320.0060+0.0060+0.03
CALCULATION

How to calculate the result

Delivered volume V = m × Z; Error % = (V − Nominal) ÷ Nominal × 100
  1. Nominal 20.000 mL, mass 19.9402 g, water at 22 °C (Z = 1.0033 mL/g)
  2. V = 19.9402 × 1.0033 = 20.0060 mL
  3. Error = (20.0060 − 20.000) ÷ 20.000 × 100 = +0.03%
ACCEPTANCE CRITERIA

Pass if

  • Systematic error within the limits in your SOP, commonly about ±0.2% of nominal for a 20 mL burette, based on ISO 8655-3 or the manufacturer's specification.
  • Random error (CV of replicates) commonly not more than 0.1%.
IF IT FAILS

What to do

  • Check for air bubbles, leaks at the valve and tubing, and a worn piston seal.
  • Repeat the test after correction; if it still fails, label the titrator 'Out of calibration' and have the burette serviced.
  • Assess the impact on results generated since the last passing check.