LimsPulse
Polarimeter · Calibration procedure

Optical rotation (quartz plate)

Measures certified quartz control plates of known rotation and compares the readings with their certified values.

Photo of a Polarimeter
Photo: Gingkoaceae, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Confirms that the angle scale of the polarimeter is accurate, so the specific optical rotation of samples is correct.

Frequency

Typically every 6 or 12 months and after repair or lamp change; many labs also do a single-plate check daily or before use. As defined in your SOP.

Learn how the Polarimeter works
YOU WILL NEED

Standards & tools

  • Certified quartz control plates with a valid certificate, e.g. one positive plate (about +34°) and one negative plate (about −17°), certified at the instrument's wavelength and temperature
BEFORE YOU START

Prerequisites

  • Instrument warmed up and temperature stable
  • Plates clean and equilibrated to the measuring temperature
  • Wavelength set to 589 nm (or the certified wavelength)
PROCEDURE

Step by step

  1. 1Zero the instrument with the empty compartment (or as the manufacturer describes).
  2. 2Insert the positive quartz plate in the marked orientation.
  3. 3Take five readings and record them.
  4. 4Repeat with the negative plate.
  5. 5Calculate the mean for each plate and the difference from the certified value.
  6. 6Check the repeatability of the readings (range or standard deviation).
RECORD

Example observation table

PlateCertified value (°)Readings (°)Mean (°)Difference (°)Result
Positive+34.62534.628, 34.622, 34.626, 34.624, 34.627+34.6254+0.0004Pass
Negative−17.312−17.312, −17.308, −17.311, −17.309, −17.310−17.3100+0.0020Pass
CALCULATION

How to calculate the result

Difference = Mean reading − Certified value
  1. Positive plate mean = (34.628 + 34.622 + 34.626 + 34.624 + 34.627) ÷ 5 = 34.6254°
  2. Difference = 34.6254 − 34.625 = +0.0004°
  3. Negative plate mean = −17.3100°; difference = −17.3100 − (−17.312) = +0.0020°
ACCEPTANCE CRITERIA

Pass if

  • Difference within the tolerance in your SOP, which is usually based on the plate certificate uncertainty and the instrument specification (commonly around ±0.01° to ±0.02°).
  • Readings repeatable within the instrument specification.
IF IT FAILS

What to do

  • Clean the plate, allow it to equilibrate and repeat.
  • Check the wavelength and the plate orientation.
  • If it still fails, label the instrument 'Out of calibration', call service and assess results since the last pass.