LimsPulse
Plate reader · Calibration procedure

Wavelength accuracy

Scans a holmium or didymium glass filter in the calibration plate and checks that the measured absorption peak positions match the certified wavelengths.

Photo of a Plate reader
Photo: Pdcook, CC BY-SA 3.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

If a monochromator is mis-set, the reader measures on the slope of an absorbance band instead of its maximum. Results then lose sensitivity and drift between instruments. Filter-based readers are checked to confirm the correct filters are installed and give the expected response.

Frequency

Typically every 6–12 months and after service, as defined in your SOP.

Learn how the Plate reader works
YOU WILL NEED

Standards & tools

  • Calibration plate with a certified holmium oxide glass (or didymium) filter
  • Certificate stating peak wavelengths (holmium glass peaks are typically near 361, 418, 453 and 536 nm; use the certificate values)
BEFORE YOU START

Prerequisites

  • Reader warmed up
  • Absorbance accuracy check passed or performed in the same session
  • Scan protocol with a step of 1 nm or smaller around each peak
PROCEDURE

Step by step

  1. 1Load the calibration plate and run a spectral scan of the holmium filter across the certified peaks (for monochromator readers).
  2. 2Use an empty (air) well or the blank position as reference.
  3. 3Find the wavelength of maximum absorbance for each certified peak.
  4. 4Calculate the difference from the certified value.
  5. 5For filter-based readers, confirm installed filter wavelengths from the supplier's check and that each filter gives the expected response on the test plate.
RECORD

Example observation table

Certified peak (nm)Measured peak (nm)Difference (nm)Result
360.9361.5+0.6Pass
418.5417.8−0.7Pass
453.2453.9+0.7Pass
536.2537.1+0.9Pass
CALCULATION

How to calculate the result

Formula
Difference (nm) = measured peak wavelength − certified peak wavelength
Worked example
  1. 1

    536.2 nm peak: 537.1 − 536.2 = +0.9 nm

  2. 2

    418.5 nm peak: 417.8 − 418.5 = −0.7 nm

  3. 3

    Largest difference +0.9 nm, within a typical ±2 nm specification

ACCEPTANCE CRITERIA

Pass if

  • Each peak typically within ±2 nm of the certified value for monochromator readers (or the manufacturer's specification)
  • Filter readers: filter centre wavelength and bandwidth as specified by the supplier
IF IT FAILS

What to do

  • Repeat with a clean plate.
  • If confirmed, label the reader out of use and request service (monochromator calibration).
  • Assess the impact on assays read near steep absorbance bands.
MORE

Other Plate reader calibrations

In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how