LimsPulse
LAF · Calibration procedure

Non-viable particle count

Counts airborne particles of ≥0.5 µm and ≥5.0 µm at defined locations in the work zone with a calibrated particle counter, to classify the unit as ISO 5 / Grade A.

Photo of a LAF
Photo: Hugh Fox, CC BY-SA 2.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Particle counts show that the air in the work zone is as clean as Grade A requires. They confirm the combined performance of the filter, the seal and the airflow pattern.

Frequency

Typically every 6 months (some sites test HEPA integrity annually), and after filter replacement, relocation or major repair, as defined in your SOP.

Learn how the LAF works
YOU WILL NEED

Standards & tools

  • Calibrated discrete airborne particle counter (e.g. 28.3 L/min) with isokinetic probe, calibrated as per ISO 21501-4
BEFORE YOU START

Prerequisites

  • Air velocity and HEPA integrity tests passed
  • Unit running for at least the stabilisation time; work zone clean and empty for 'at rest' testing
  • Number of sampling locations and sample volume calculated as per ISO 14644-1 and your SOP
PROCEDURE

Step by step

  1. 1Place the isokinetic probe at working height, facing into the airflow, at each location.
  2. 2Sample at least the minimum volume required at each location (sites commonly take 1 m³ for Grade A classification, or as per your SOP).
  3. 3Record counts for ≥0.5 µm and ≥5.0 µm and convert them to particles per m³.
  4. 4Repeat in operation (with a simulated activity) if the SOP requires 'in operation' classification.
  5. 5Compare each location with the limits.
RECORD

Example observation table

LocationVolume sampled (L)Count ≥0.5 µm≥0.5 µm per m³≥5.0 µm per m³
L1 (left)28.362120
L2 (centre)28.341410
L3 (right)28.393180
CALCULATION

How to calculate the result

Particles per m³ = Count ÷ Volume sampled (m³) (1 m³ = 1000 L)
  1. Location L3: 9 particles ≥0.5 µm in 28.3 L = 0.0283 m³
  2. Concentration = 9 ÷ 0.0283 = 318 particles/m³
  3. Limit for ISO 5 / Grade A at ≥0.5 µm is 3,520 /m³: Pass
ACCEPTANCE CRITERIA

Pass if

  • ≥0.5 µm: not more than 3,520 particles/m³ at every location (ISO 5 / Grade A, at rest and in operation).
  • ≥5.0 µm: in line with Annex 1 expectations for Grade A (no limit for ISO classification; Annex 1 gives 29 /m³ in operation), as defined in your SOP.
IF IT FAILS

What to do

  • Check the counter calibration, probe position and sampling technique, then repeat the count.
  • Clean the unit and check velocity and HEPA integrity.
  • Investigate room conditions and operator activity for 'in operation' failures.
  • Do not use the unit for aseptic work until it passes.