LAF · Calibration procedure

HEPA filter integrity

Challenges the HEPA filter with an oil aerosol (PAO or DOP) upstream and scans the whole downstream face and seal with a photometer probe to detect leaks.

Photo of a LAF
Photo: Hugh Fox, CC BY-SA 2.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

A pinhole or seal leak lets unfiltered air into the Grade A zone even when average velocity and particle counts look acceptable. The integrity scan detects such leaks so they can be repaired before they contaminate work.

Frequency

Typically every 6 months (some sites test HEPA integrity annually), and after filter replacement, relocation or major repair, as defined in your SOP.

Learn how the LAF works
YOU WILL NEED

Standards & tools

  • Aerosol generator producing a polydisperse PAO (poly-alpha-olefin) or DOP aerosol
  • Calibrated aerosol photometer
  • Alternatively, a discrete particle counter method as described in ISO 14644-3
BEFORE YOU START

Prerequisites

  • Air velocity checked and within limits
  • Upstream sample port accessible; diffuser or grille removed if needed
  • Unit cleaned before the test and plan to clean after it (aerosol leaves an oil residue)
PROCEDURE

Step by step

  1. 1Introduce the aerosol upstream of the HEPA filter and measure the upstream concentration (typically 10–100 µg/L, often about 20 µg/L); set this as 100% on the photometer.
  2. 2Scan the entire downstream face of the filter with the probe about 2–3 cm from the surface, in slightly overlapping strokes, at no more than about 5 cm/s.
  3. 3Scan the full perimeter, including the gasket and the frame-to-housing seal.
  4. 4Where the reading exceeds the limit, hold the probe still to confirm and mark the leak location.
  5. 5Record the maximum penetration found and recheck the upstream concentration at the end.
RECORD

Example observation table

LocationUpstream concentration (µg/L)Max. downstream reading (% of upstream)Result
Filter face, zone 120.00.002Pass
Filter face, zone 220.00.003Pass
Filter face, zone 320.00.001Pass
Perimeter seal20.00.002Pass
CALCULATION

How to calculate the result

Penetration (%) = Downstream concentration ÷ Upstream concentration × 100
  1. Upstream = 20.0 µg/L; highest downstream reading in zone 2 = 0.0006 µg/L
  2. Penetration = 0.0006 ÷ 20.0 × 100 = 0.003%
  3. Below a typical leak limit of 0.01%: Pass
ACCEPTANCE CRITERIA

Pass if

  • No point with penetration greater than 0.01% of the upstream concentration (typical leak criterion), as defined in your SOP.
  • Upstream concentration within the range required by the method at the start and end of the test.
IF IT FAILS

What to do

  • Stop using the unit.
  • Repair small face leaks with an approved sealant if the manufacturer allows and the total repaired area is within the permitted limit; otherwise replace the filter.
  • Reseat or replace a leaking gasket.
  • Repeat the integrity test, then re-check velocity and particle count before release.