LimsPulse
Viscosity bath · Calibration procedure

Viscometer constant verification

Measures the flow time of a certified viscosity standard in a routine viscometer and compares the calculated constant with the certified constant.

Illustration of Viscosity bath
PURPOSE

Why this is checked

Each glass capillary viscometer has its own constant C (mm²/s²). Wear, deposits or damage can change it, and a wrong constant gives a proportional error in every result. Running a certified viscosity standard checks the constant, and also checks the bath temperature, timing and technique together.

Frequency

Typically annually for each viscometer, when a new viscometer is put into service, after a check standard fails, or per SOP.

Learn how the Viscosity bath works
YOU WILL NEED

Standards & tools

  • Certified kinematic viscosity standard (oil) with a value certified at the test temperature, chosen so the flow time is at least 200 s
  • Calibrated bath and reference thermometer
  • Timer verified against a reference time signal
BEFORE YOU START

Prerequisites

  • Bath temperature verified
  • Viscometer clean and dry
  • Certificate value for the standard at the exact test temperature
PROCEDURE

Step by step

  1. 1Charge the viscometer with the standard and equilibrate in the bath for about 30 minutes.
  2. 2Measure two flow times and check they agree within the determinability.
  3. 3Calculate the mean flow time.
  4. 4Calculate the determined constant: C = ν(standard) ÷ mean flow time.
  5. 5Compare with the certified constant of the viscometer and calculate the percentage difference.
RECORD

Example observation table

ItemValue
Standard viscosity at 40.00 °C (certificate)61.23 mm²/s
Flow time 1408.6 s
Flow time 2408.2 s
Certified viscometer constant0.1500 mm²/s²
CALCULATION

How to calculate the result

Formula
C(determined) = ν(standard) ÷ t(mean); Difference (%) = (C(determined) − C(certified)) ÷ C(certified) × 100
Worked example
  1. 1
    Mean flow time = (408.6 + 408.2) ÷ 2 = 408.4 s.
  2. 2
    C(determined) = 61.23 ÷ 408.4 = 0.14993 mm²/s².
  3. 3
    Difference = (0.14993 − 0.1500) ÷ 0.1500 × 100 = −0.05%.
  4. 4

    Within an illustrative SOP limit of ±0.2% — the certified constant is confirmed.

ACCEPTANCE CRITERIA

Pass if

  • Determined constant agrees with the certified constant within the limit in your SOP (commonly a few tenths of a per cent), or the measured viscosity of the standard is within the certificate tolerance.
  • Two flow times agree within the determinability.
IF IT FAILS

What to do

  • Clean the viscometer thoroughly (deposits change the constant), re-check verticality and bath temperature, and repeat.
  • If it still fails, send the viscometer for recalibration or withdraw it.
  • Assess results obtained with that viscometer since the last good check.
MORE

Other Viscosity bath calibrations

In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.