Viscometer constant verification
Measures the flow time of a certified viscosity standard in a routine viscometer and compares the calculated constant with the certified constant.
Why this is checked
Each glass capillary viscometer has its own constant C (mm²/s²). Wear, deposits or damage can change it, and a wrong constant gives a proportional error in every result. Running a certified viscosity standard checks the constant, and also checks the bath temperature, timing and technique together.
Typically annually for each viscometer, when a new viscometer is put into service, after a check standard fails, or per SOP.
Learn how the Viscosity bath worksStandards & tools
- Certified kinematic viscosity standard (oil) with a value certified at the test temperature, chosen so the flow time is at least 200 s
- Calibrated bath and reference thermometer
- Timer verified against a reference time signal
Prerequisites
- Bath temperature verified
- Viscometer clean and dry
- Certificate value for the standard at the exact test temperature
Step by step
- 1Charge the viscometer with the standard and equilibrate in the bath for about 30 minutes.
- 2Measure two flow times and check they agree within the determinability.
- 3Calculate the mean flow time.
- 4Calculate the determined constant: C = ν(standard) ÷ mean flow time.
- 5Compare with the certified constant of the viscometer and calculate the percentage difference.
Example observation table
| Item | Value |
|---|---|
| Standard viscosity at 40.00 °C (certificate) | 61.23 mm²/s |
| Flow time 1 | 408.6 s |
| Flow time 2 | 408.2 s |
| Certified viscometer constant | 0.1500 mm²/s² |
How to calculate the result
- 1Mean flow time = (408.6 + 408.2) ÷ 2 = 408.4 s.
- 2C(determined) = 61.23 ÷ 408.4 = 0.14993 mm²/s².
- 3Difference = (0.14993 − 0.1500) ÷ 0.1500 × 100 = −0.05%.
- 4
Within an illustrative SOP limit of ±0.2% — the certified constant is confirmed.
Pass if
- Determined constant agrees with the certified constant within the limit in your SOP (commonly a few tenths of a per cent), or the measured viscosity of the standard is within the certificate tolerance.
- Two flow times agree within the determinability.
What to do
- Clean the viscometer thoroughly (deposits change the constant), re-check verticality and bath temperature, and repeat.
- If it still fails, send the viscometer for recalibration or withdraw it.
- Assess results obtained with that viscometer since the last good check.