LimsPulse
IC · Calibration procedure

Conductivity detector linearity

Checks that the suppressed conductivity response rises in proportion to ion concentration, using a series of chloride (or mixed-anion) standards.

Photo of a IC
Photo: Datamax, Public domain (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

IC results are calculated from a calibration curve. If the detector and suppressor response is not linear over the working range, results at low or high concentrations will be biased.

Frequency

Typically every 6 months, and after major repair or relocation, as defined in your calibration SOP.

Learn how the IC works
YOU WILL NEED

Standards & tools

  • Certified chloride stock standard (e.g. 1000 mg/L, traceable)
  • Deionised water (18.2 MΩ·cm) for dilutions
  • Optionally, a potassium chloride conductivity standard (0.001 mol/L KCl is about 147 µS/cm at 25 °C) to check the cell reading
BEFORE YOU START

Prerequisites

  • Equilibrate the system with the routine anion column, eluent and suppressor until background conductivity is stable
  • Inject a water blank and confirm there is no significant chloride peak
PROCEDURE

Step by step

  1. 1Prepare at least five standards across the working range, e.g. 0.5, 1, 2, 5 and 10 mg/L chloride.
  2. 2Inject each standard (in duplicate if your SOP requires) from lowest to highest.
  3. 3Record the chloride peak area at each level.
  4. 4Plot peak area against concentration and calculate the slope, intercept and correlation coefficient (r).
RECORD

Example observation table

Chloride (mg/L)Peak area (µS·min)
0.50.061
1.00.124
2.00.246
5.00.618
10.01.229
CALCULATION

How to calculate the result

Least-squares line: Area = slope × Concentration + intercept; r = Σ(x − x̄)(y − ȳ) ÷ √[Σ(x − x̄)² × Σ(y − ȳ)²]
  1. Slope = 0.1229 µS·min per mg/L; intercept = 0.0007 µS·min
  2. r = 0.99999 (r² = 0.99999)
  3. Result meets r ≥ 0.999
Use the Standard curve (linear regression) calculator
ACCEPTANCE CRITERIA

Pass if

  • Correlation coefficient r typically not less than 0.999 over the range; intercept small relative to the lowest standard, as per your SOP.
IF IT FAILS

What to do

  • Check the suppressor (exhausted or damaged suppressors cause non-linear response) and background conductivity.
  • Check standard preparation and water quality.
  • Check cell temperature control; service the detector if needed.
  • Repeat after correction; label out of calibration until it passes.
MORE

Other IC calibrations

In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how