LimsPulse
ICP-MS · Calibration procedure

Tuning & mass calibration

Uses a tuning solution of light, medium and heavy elements to check and adjust mass accuracy, peak width, sensitivity and the oxide and doubly charged ion ratios.

Photo of a ICP-MS
Photo: Cyberwork 95, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Ions must be measured at the right mass with enough resolution and sensitivity, and the plasma must produce few interfering oxide and doubly charged ions. Tuning confirms the instrument is fit before any elemental impurities analysis.

Frequency

Typically tuning check each day of use; full mass calibration and detection limit calibration every 6 or 12 months and after cone, lens or detector service, as defined in your calibration SOP.

Learn how the ICP-MS works
YOU WILL NEED

Standards & tools

  • Manufacturer's tuning solution (e.g. Li, Co, Y, Ce, Tl at about 1 µg/L in 2% nitric acid), traceable
  • 2% nitric acid blank
BEFORE YOU START

Prerequisites

  • Vacuum in normal range
  • Plasma lit and warmed up (often 20–30 minutes)
  • Clean cones and fresh pump tubing
PROCEDURE

Step by step

  1. 1Aspirate the tuning solution.
  2. 2Run the auto-tune or performance report for the mode used (no gas and helium mode).
  3. 3Record sensitivity (cps) for low, mid and high mass, background, oxide ratio (CeO⁺/Ce⁺) and doubly charged ratio (Ce²⁺/Ce⁺).
  4. 4Check measured peak centres and peak widths for each mass; run a mass calibration if any are outside limits.
  5. 5Save the tune and print or store the report.
RECORD

Example observation table

ParameterMeasuredExample limit
⁷Li mass / peak width (10% height)7.00 / 0.72 amu±0.1 amu / 0.65–0.85 amu
⁸⁹Y mass / peak width88.95 / 0.74 amu±0.1 amu / 0.65–0.85 amu
²⁰⁵Tl mass / peak width205.03 / 0.75 amu±0.1 amu / 0.65–0.85 amu
¹⁴⁰Ce⁺ counts98,000 cpsAs specified
¹⁵⁶CeO⁺ counts1,250 cps—
⁷⁰Ce²⁺ counts1,860 cps—
CALCULATION

How to calculate the result

Oxide ratio (%) = CeO⁺ ÷ Ce⁺ × 100; Doubly charged ratio (%) = Ce²⁺ ÷ Ce⁺ × 100
  1. Oxide ratio = 1,250 ÷ 98,000 × 100 = 1.28%
  2. Doubly charged ratio = 1,860 ÷ 98,000 × 100 = 1.90%
  3. Both below the example limits of 2–3%: passes
ACCEPTANCE CRITERIA

Pass if

  • Mass accuracy typically within ±0.1 amu and peak width (10% height) within the manufacturer's range (often about 0.65–0.85 amu).
  • Oxide (CeO⁺/Ce⁺) and doubly charged (Ce²⁺/Ce⁺) ratios typically below about 2–3%, and sensitivity above the manufacturer's minimum. Use the limits in your SOP.
IF IT FAILS

What to do

  • Re-run auto-tune; check nebuliser gas flow, torch position and spray chamber temperature.
  • Clean or replace cones; check for interface leaks.
  • Run a full mass calibration; call service if it still fails.
  • Do not analyse samples until the tune passes.
MORE

Other ICP-MS calibrations

In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how