LimsPulse
Hot air oven · Calibration procedure

Temperature mapping

Measures the temperature at several positions in the chamber at the same time to show how uniform the oven is.

Photo of a Hot air oven
Photo: Karelj, Public domain (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Ovens have warmer and cooler zones. Mapping shows which shelf positions stay within limits, so samples are only placed where the temperature is correct.

Frequency

At OQ/PQ, then typically annually and after major repair, relocation or change of load pattern, as defined in your SOP.

Learn how the Hot air oven works
YOU WILL NEED

Standards & tools

  • Multi-channel data logger with at least 9 calibrated probes (for a small oven), traceable to national standards
BEFORE YOU START

Prerequisites

  • Oven empty (or with a defined typical load for PQ)
  • Probes calibrated before and verified after the study
  • Probe positions drawn on a layout diagram
PROCEDURE

Step by step

  1. 1Place probes at the eight corners of the usable space (about 5 cm from the walls) and one at the centre.
  2. 2Set the oven to the working temperature (for example 105 °C).
  3. 3After stabilisation, record all probes every minute for at least 30 minutes.
  4. 4Calculate the mean at each position and the overall minimum, maximum and spread.
  5. 5Mark positions that fail as 'not for use' on the oven layout.
RECORD

Example observation table

ProbePositionMean over 30 min (°C)
1Top front left104.2
2Top front right105.6
3Top back left104.8
4Top back right105.9
5Centre105.1
6Bottom front left104.4
7Bottom front right106.1
8Bottom back left104.7
9Bottom back right105.3
CALCULATION

How to calculate the result

Uniformity (spread) = Highest mean − Lowest mean; Deviation = Probe mean − Set temperature
  1. Highest = 106.1 °C (probe 7), lowest = 104.2 °C (probe 1)
  2. Spread = 106.1 − 104.2 = 1.9 °C; overall mean = 105.1 °C
  3. Largest deviation = 106.1 − 105 = +1.1 °C (within ±2 °C)
ACCEPTANCE CRITERIA

Pass if

  • Every probe within ±2 °C of the set temperature throughout the study (typical; use the limit in your SOP).
  • Spread between the hottest and coldest positions within the SOP limit (for example not more than 4 °C).
IF IT FAILS

What to do

  • Identify hot or cold spots and restrict use of those positions, or service the fan, heaters or door seal.
  • Repeat the mapping after correction.
  • Assess results generated in positions that were outside the limit.
MORE

Other Hot air oven calibrations

In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how