GC · Calibration procedure
Oven temperature accuracy
Checks that the column oven reaches and holds the set temperature, using a calibrated temperature probe placed near the column.

PURPOSE
Why this is checked
Retention times in GC depend strongly on oven temperature. An oven that reads high or low shifts retention, changes resolution between close peaks and makes methods non-transferable.
Frequency
Typically every 6 or 12 months, and after major repair or relocation, as defined in your calibration SOP.
Learn how the GC worksYOU WILL NEED
Standards & tools
- Calibrated digital thermometer with a thermocouple or Pt100 probe covering at least 40–300 °C, traceable to national standards
BEFORE YOU START
Prerequisites
- Fix the probe in the centre of the oven near the column hanger, not touching the walls or the column
- Close the oven door with the probe lead routed through the door seal or a port
- Switch off the detector flame if the probe lead passes near it
PROCEDURE
Step by step
- 1Set the oven to the first set point (e.g. 40 °C) and wait at least 10–15 minutes after the instrument shows it is ready.
- 2Record the probe reading (take the mean of a few readings over 1–2 minutes).
- 3Repeat at set points covering the range your methods use, for example 100, 200 and 300 °C.
- 4Optionally record readings every 30 s for 10 minutes at one set point to check stability.
RECORD
Example observation table
| Set (°C) | Observed (°C) | Difference (°C) |
|---|---|---|
| 40 | 40.3 | +0.3 |
| 100 | 100.6 | +0.6 |
| 200 | 199.2 | −0.8 |
| 300 | 298.9 | −1.1 |
CALCULATION
How to calculate the result
Difference = Observed − Set
- Set 300 °C, observed 298.9 °C
- Difference = 298.9 − 300 = −1.1 °C
- Within a typical ±2 °C limit, so it passes
ACCEPTANCE CRITERIA
Pass if
- Typically within ±2 °C of the set value at each point (some SOPs use ±1 °C or ±1% of the set value). Use the limit in your SOP.
- Stability, if tested: typically within ±0.5 °C over 10 minutes.
IF IT FAILS
What to do
- Label the instrument 'Out of calibration' and stop using it.
- Check probe position and repeat once to rule out a measurement error.
- Get the oven sensor or controller serviced and adjusted by an engineer.
- Assess the impact on results generated since the last passing calibration.