GC · Calibration procedure

Injector precision

Checks that the autosampler and inlet deliver the same amount of sample each time, by calculating the %RSD of peak areas from replicate injections.

Photo of a GC
Photo: Sarka Na kopci, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Every quantitative GC result assumes the injection is reproducible. Poor injector precision shows up as variable peak areas and failed system suitability.

Frequency

Typically every 6 or 12 months, and after major repair or relocation, as defined in your calibration SOP.

Learn how the GC works
YOU WILL NEED

Standards & tools

  • Certified test mixture or a solution of a stable compound such as n-dodecane or n-tetradecane in hexane (concentration as per your SOP)
BEFORE YOU START

Prerequisites

  • New septum and clean liner fitted
  • Suitable column (e.g. 5% phenyl polysiloxane) and FID conditions set as in the calibration SOP
  • Baseline stable after a blank injection
PROCEDURE

Step by step

  1. 1Inject the test solution six times in split mode at a fixed volume (e.g. 1 µL).
  2. 2Integrate the main peak in each chromatogram with the same integration parameters.
  3. 3Calculate the mean, standard deviation and %RSD of the peak areas.
  4. 4Optionally calculate the %RSD of retention time as well.
RECORD

Example observation table

InjectionRetention time (min)Peak area (pA·s)
18.412245,310
28.413243,880
38.411246,920
48.412244,150
58.414247,360
68.412245,020
CALCULATION

How to calculate the result

%RSD = (Standard deviation ÷ Mean) × 100
  1. Mean area = 245,440 pA·s
  2. Standard deviation (n − 1) = 1,426
  3. %RSD = 1,426 ÷ 245,440 × 100 = 0.58%
Use the Mean, SD and %RSD calculator
ACCEPTANCE CRITERIA

Pass if

  • Typically %RSD of peak area not more than 2.0% for FID with split injection (n = 6), or the limit in your SOP.
IF IT FAILS

What to do

  • Replace the septum and liner and check the syringe for sticking or air bubbles.
  • Check the split vent flow and inlet for leaks.
  • Service the autosampler and repeat the test.
  • Do not use the instrument until it passes.