Disintegration tester · Calibration procedure

Temperature

Checks the immersion fluid temperature against a calibrated thermometer.

Illustration of Disintegration tester
PURPOSE

Why this is checked

Disintegration is faster in warm fluid; the test is only valid at the stated temperature.

Frequency

Typically every 3 to 6 months; the fluid temperature is also checked at each test.

Learn how the Disintegration tester works
YOU WILL NEED

Standards & tools

  • Calibrated digital thermometer, 0.1 °C resolution
BEFORE YOU START

Prerequisites

  • Beakers filled with water
  • Bath stabilised at the set temperature
PROCEDURE

Step by step

  1. 1Set the controller to 37.0 °C and allow at least 20–30 minutes to stabilise.
  2. 2Place the calibrated probe in each beaker near the centre and record the reading and the displayed value.
  3. 3Repeat for each station.
RECORD

Example observation table

StationSet (°C)Displayed (°C)Observed (°C)Difference (°C)
A37.037.037.2+0.2
B37.037.036.8−0.2
CALCULATION

How to calculate the result

Difference = Observed − Set
  1. Station A: 37.2 − 37.0 = +0.2 °C (within ±2 °C)
ACCEPTANCE CRITERIA

Pass if

  • Immersion fluid typically 37 ± 2 °C.
  • Many SOPs also require the display to agree with the reference within a tighter tolerance, such as ±0.5 °C.
IF IT FAILS

What to do

  • Check bath level and heater
  • Adjust or replace the sensor
  • Do not use until it passes