LimsPulse
BSC · Calibration procedure

HEPA filter integrity

Challenges the supply and exhaust HEPA filters of the cabinet with an aerosol (PAO/DOP) and scans the downstream side with a photometer to find leaks in the filter media or seals.

Photo of a BSC
Photo: The wub, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

A leak in the supply filter lets unfiltered air onto the work, and a leak in the exhaust filter lets microorganisms escape into the room. Filter integrity is therefore tested on both filters at each certification.

Frequency

At installation, then typically every 12 months (some sites every 6 months), and after filter change, relocation or major repair, as defined in your SOP.

Learn how the BSC works
YOU WILL NEED

Standards & tools

  • Aerosol generator (PAO or equivalent oil aerosol)
  • Calibrated aerosol photometer with scanning probe
  • Alternatively, a discrete particle counter method where the standard allows it
BEFORE YOU START

Prerequisites

  • Inflow and downflow velocities within limits
  • Cabinet decontaminated if the filter housing must be opened
  • Upstream sample port accessible
PROCEDURE

Step by step

  1. 1Introduce the aerosol into the cabinet (typically at the work surface or into the plenum, as per the manufacturer) and measure the upstream concentration; set it as 100% on the photometer.
  2. 2Supply filter: scan the whole downstream face and its perimeter with the probe about 2–3 cm from the filter, in overlapping strokes at no more than about 5 cm/s.
  3. 3Exhaust filter: scan the downstream face if accessible; if it cannot be scanned (e.g. ducted), take a single-point reading downstream as the standard allows.
  4. 4Hold the probe still on any point above the limit to confirm and mark it.
  5. 5Recheck the upstream concentration at the end and record the maximum penetration for each filter.
RECORD

Example observation table

Filter / zoneUpstream (µg/L)Max. downstream (% of upstream)Result
Supply filter, left third15.00.004Pass
Supply filter, centre third15.00.008Pass
Supply filter, right third15.00.003Pass
Supply filter perimeter seal15.00.002Pass
Exhaust filter (scan)15.00.002Pass
CALCULATION

How to calculate the result

Penetration (%) = Downstream concentration ÷ Upstream concentration × 100
  1. Upstream = 15.0 µg/L; highest downstream reading on the supply filter centre = 0.0012 µg/L
  2. Penetration = 0.0012 ÷ 15.0 × 100 = 0.008%
  3. Not more than the typical scan limit of 0.01%: Pass
ACCEPTANCE CRITERIA

Pass if

  • Scanned filters: no point with penetration above 0.01% of upstream concentration (typical criterion in NSF/ANSI 49 and common SOPs).
  • Filters that cannot be scanned: typically not more than 0.005% at a single downstream point (NSF/ANSI 49 approach), or as defined in your SOP.
  • Upstream concentration within the range required by the procedure at start and end.
IF IT FAILS

What to do

  • Stop using the cabinet.
  • Repair a small face leak with approved sealant if the manufacturer allows and the repaired area is within limits; otherwise decontaminate and replace the filter.
  • Reseat or replace the gasket for seal leaks.
  • Repeat integrity, velocity and smoke tests before release.
MORE

Other BSC calibrations

In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how