Polymers, plastics & coatings · Moulded sheet, plaque or finished part of rubber or plastic

Shore hardness of rubbers and plastics

A calibrated Shore durometer is pressed onto a flat specimen of the required thickness. The reading after a set time is recorded at several points, and the median or mean is reported with the scale and dwell time.

  • Chemicals & materials
  • Durometer indentation
  • Durometer

Reference: Typical method based on ISO 868 (plastics and ebonite), ISO 48-4 (rubber) and ASTM D2240.

PRINCIPLE

Why this method works

The durometer indentor penetrates the specimen against a calibrated spring. Penetration of 0 to 2.5 mm is shown as 100 to 0 Shore units. Shore A suits soft rubbers and elastomers; Shore D suits hard rubbers and rigid plastics. Because polymers creep, the reading decreases with time, so the dwell time is fixed.

Instruments used

SAMPLE PREPARATION

Preparing the sample

  1. 01

    Prepare

    Cut or mould flat specimens of at least the minimum thickness.

  2. 02

    Condition

    Condition at 23 °C (and controlled humidity if required) for the required time.

  3. 03

    Measure

    Apply the durometer and read after the stated time.

  4. 04

    Repeat

    Take at least five readings at different points.

Procedure

  1. 1Use specimens with a flat, smooth surface and thickness typically at least 4 mm (ISO 868) or 6 mm (ASTM D2240). Thinner sheets may be plied if the standard allows.
  2. 2Condition specimens at 23 ± 2 °C for the time in your SOP (often at least 1 h, longer for freshly moulded parts).
  3. 3Check the durometer zero and verify it on a reference block.
  4. 4Place the specimen on a hard, flat surface. Apply the durometer vertically (preferably with an operating stand) at least about 9–12 mm from any edge.
  5. 5Read the scale after the dwell time: 15 s for ISO 868, within 1 s or a stated time for ASTM D2240, or as the product specification states.
  6. 6Take at least five readings at points at least about 6 mm apart; record each one.

Standard preparation

  1. 1Reference rubber blocks of known hardness are used for daily checks of the durometer.

Notes for analysts

  • Shore A and Shore D values cannot be converted exactly into each other or into IRHD or Rockwell scales.
  • Readings on curved or thin finished parts are usually lower than on standard flat specimens; agree the test with the customer.
  • Write results as, for example, 'Shore A 73 (15 s)' so the scale and dwell time are clear.
TEST CONDITIONS

Method parameters (typical)

ScaleShore A for soft materials (readings about 10–90); Shore D for hard materials (readings above about 20)
Dwell timeAs per standard: for example 15 s (ISO 868) or 1 s (ASTM D2240 instantaneous); always state it
Specimen thicknessTypically ≥ 4 mm (ISO 868) or ≥ 6 mm (ASTM D2240)
Temperature23 ± 2 °C
ReadingsAt least five per specimen
BEFORE YOU REPORT

System suitability & sequence

ParameterLimit
Reference blockDurometer reads the reference block within its stated tolerance
ZeroReads 0 with indentor free

Typical sequence

  1. 1Condition specimens
  2. 2Check durometer
  3. 3Measure five points
  4. 4Report median/mean with scale and time
CALCULATION

Calculation with a worked example

Formula
Result = median (or mean) of n readings, as the standard specifies

Where

n
Number of readings (typically 5)

Given

Readings (Shore A, 15 s)
72, 74, 73, 71, 73

Working

  1. 1

    Sorted readings: 71, 72, 73, 73, 74 → median = 73

  2. 2
    Mean = (72 + 74 + 73 + 71 + 73) ÷ 5 = 363 ÷ 5 = 72.6

Hardness = 73 Shore A (15 s), median of five readings

ACCEPTANCE CRITERIA

What result passes

  • Compare with the drawing or material specification, usually nominal ± 5 Shore units for rubber compounds (check your specification).
METHOD VALIDATION

Validation parameters (ICH Q2(R2))

ParameterTypical requirement
PrecisionRepeatability between readings and between operators; stand-mounted operation improves it