Polymers, plastics & coatings · Paint, varnish or coating film applied to a flat test panel; plastic plaques

Gloss of paints and coatings

A coating is applied to a flat panel at a set film thickness and cured. Gloss is measured with a calibrated gloss meter at 60°, or at 20° or 85° depending on the gloss level, and the mean of several readings is reported in gloss units.

  • Chemicals & materials
  • Specular reflectance (gloss meter)
  • Gloss meter

Reference: Typical method based on ISO 2813 and ASTM D523.

PRINCIPLE

Why this method works

Smooth, glossy surfaces reflect more light in the mirror direction. The gloss meter compares specular reflectance with a black glass standard assigned 100 GU. The 60° geometry is used for most coatings; 20° separates high-gloss finishes, and 85° separates matt finishes.

Instruments used

SAMPLE PREPARATION

Preparing the sample

  1. 01

    Apply

    Draw down the coating on a flat glass or metal panel with a film applicator at the set wet thickness.

  2. 02

    Cure

    Dry or stove the film and condition it as the method states.

  3. 03

    Calibrate

    Standardise the gloss meter on its calibration tile.

  4. 04

    Measure

    Measure at 60° first, then choose the geometry.

  5. 05

    Report

    Report the mean, range and geometry.

Procedure

  1. 1Apply the coating to a clean, flat, rigid panel (for example glass or a standard metal panel) using a film applicator of the stated gap.
  2. 2Dry or stove the film and condition it, typically for at least 16 h at 23 °C and 50% RH, or as your method states.
  3. 3Standardise the gloss meter on its high-gloss tile and check a lower-gloss tile.
  4. 4Measure at 60°. If the reading is above about 70 GU, also measure at 20°; if below about 10 GU, measure at 85°.
  5. 5Take at least three readings at different places on the panel, and in a set direction relative to the application direction.
  6. 6Calculate the mean and record the range.

Standard preparation

  1. 1Keep a retained reference panel of the product standard to measure alongside each batch.

Notes for analysts

  • Always state the geometry: '85 GU' means little unless it says 60°, 20° or 85°.
  • Film thickness, substrate, drying and pigment dispersion all affect gloss; prepare batch and standard panels the same way.
  • Gloss retention (%) after weathering = gloss after exposure ÷ initial gloss × 100.
TEST CONDITIONS

Method parameters (typical)

Geometry60° (general); 20° for high gloss (60° reading > about 70 GU); 85° for low gloss (60° reading < about 10 GU)
SubstrateFlat glass or metal panel; black glass for some clear coatings
ConditioningAs per method, typically 23 °C / 50% RH
ReadingsAt least three per panel
BEFORE YOU REPORT

System suitability & sequence

ParameterLimit
CalibrationMeter standardised on its tile and check tile within tolerance
Reading spreadRange of readings on one panel within the limit in your SOP

Typical sequence

  1. 1Prepare and cure panels
  2. 2Calibrate meter
  3. 3Measure reference panel
  4. 4Measure batch panel(s)
ACCEPTANCE CRITERIA

What result passes

  • Compare with the product specification, for example 'gloss at 60° = 85 ± 5 GU' or a range for matt, satin or gloss grades. Typical descriptions: matt below about 10 GU, satin/semi-gloss roughly 10–70 GU, gloss above about 70 GU at 60° (definitions vary by industry).
METHOD VALIDATION

Validation parameters (ICH Q2(R2))

ParameterTypical requirement
PrecisionRepeatability and reproducibility as given in ISO 2813 / ASTM D523 precision statements