Leak / Vacuum Tester

Checks that packs and containers are sealed, by pulling a vacuum around them and looking for dye ingress, bubbles or a rise in chamber pressure. Covers simple blister leak tests and deterministic vacuum-decay CCI testing.

  • Pharmaceutical
  • Medical devices
  • Food packaging
  • Nutraceutical
Follow your site SOP
This is a typical operating sequence for learning. Your instrument model and approved SOP take precedence.
STAGE 1

Before you start

  1. 1Check the gauge and tester are within calibration.
  2. 2For dye tests, prepare fresh methylene blue solution as per the SOP.
  3. 3For vacuum decay, select the validated recipe for the container type and run the positive and negative controls.
  • Inspect the glass chamber for cracks before applying vacuum; a cracked vessel can implode.
STAGE 2

Dye ingress (blister) test

  1. 1Place the strips in the chamber and cover them with dye solution, using the hold-down plate.
  2. 2Close the lid and apply the set vacuum (as per SOP) for the set time.
  3. 3Release the vacuum slowly and leave the strips in the dye for the stated time.
  4. 4Remove, rinse and dry the strips, then open each pocket and check for blue dye.
STAGE 3

Vacuum decay test

  1. 1Place the container in the test chamber and close it.
  2. 2Start the cycle; the tester evacuates, stabilises and measures the pressure rise.
  3. 3Read the pass/fail result and record it.
  4. 4Re-run the positive control at the end of the session to show the system still detects leaks.
  • Containers wet on the outside can give false failures; dry them first.