Water bath · Calibration procedure

Temperature accuracy and uniformity

Measures the water temperature at several positions with a calibrated reference thermometer to check how close the bath is to its set point and how even it is.

Photo of a Water bath
Photo: Athikhun.suw, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Tests that depend on temperature (e.g. keeping agar molten or incubating a reaction) need the bath to be at the right temperature everywhere samples are placed.

Frequency

Typically every 6 or 12 months at the set points used in methods, as defined in your calibration SOP.

Learn how the Water bath works
YOU WILL NEED

Standards & tools

  • Calibrated reference thermometer or multi-channel probe logger, traceable to national standards
BEFORE YOU START

Prerequisites

  • Bath filled to the normal level and lid in place
  • Bath stable at the set point for at least 30 minutes
PROCEDURE

Step by step

  1. 1Choose the set points used in methods (e.g. 37 °C and 45 °C).
  2. 2Place the reference probe at five positions: four corners and the centre, at sample depth.
  3. 3Allow each reading to stabilise, then record the reference reading and the bath display.
  4. 4Calculate the mean (accuracy) and the spread between positions (uniformity).
RECORD

Example observation table

PositionReference (°C)Display (°C)
Front left36.837.0
Front right37.137.0
Rear left36.937.0
Rear right37.337.0
Centre37.037.0
CALCULATION

How to calculate the result

Accuracy = Mean reference − Set point; Uniformity = Maximum − Minimum reference reading
  1. Set point 37.0 °C
  2. Mean = (36.8 + 37.1 + 36.9 + 37.3 + 37.0) ÷ 5 = 185.1 ÷ 5 = 37.02 °C
  3. Accuracy = 37.02 − 37.0 = +0.02 °C
  4. Uniformity = 37.3 − 36.8 = 0.5 °C
ACCEPTANCE CRITERIA

Pass if

  • Each position typically within ±1 °C of the set point (tighter, e.g. ±0.5 °C, where a method requires).
  • Uniformity as defined in your SOP.
IF IT FAILS

What to do

  • Label the bath and stop using it for temperature-critical tests.
  • Check water level, circulation and controller; service and repeat.