LimsPulse
aw meter · Calibration procedure

Verification with salt standards

Reads salt solutions of known water activity and checks the meter reads each one within tolerance.

Illustration of aw meter
PURPOSE

Why this is checked

Saturated and unsaturated salt solutions have stable, well-known water activities at a given temperature. Reading them confirms the sensor and temperature measurement are accurate across the range used for samples.

Frequency

Typically daily or before each session with one or two standards near the sample range, and a wider multi-point check monthly, as defined in your SOP.

Learn how the aw meter works
YOU WILL NEED

Standards & tools

  • Sealed verification standards from the instrument supplier (for example 0.250, 0.760 and 0.984 aw at 25 °C)
  • Or saturated salt slurries at 25 °C: lithium chloride about 0.113, magnesium chloride about 0.328, sodium chloride about 0.753, potassium chloride about 0.843, potassium sulphate about 0.973
  • Distilled water (aw 1.000) as a top-end check if the manual recommends it
BEFORE YOU START

Prerequisites

  • Instrument at its set temperature (commonly 25 °C) and switched on long enough to stabilise
  • Standards in date, sealed and at room temperature
  • Chamber clean and dry
PROCEDURE

Step by step

  1. 1Open a standard vial and pour it into a clean sample cup (or use the cup supplied).
  2. 2Load it into the chamber and start the reading.
  3. 3Record the aw and temperature once stable.
  4. 4Repeat with the other standards, from low aw to high aw.
  5. 5Compare each reading with the standard value at that temperature.
RECORD

Example observation table

StandardExpected aw (25 °C)ReadingDifference
LiCl 13.41 mol/kg0.2500.252+0.002
NaCl 6.0 mol/kg0.7600.762+0.002
KCl 0.5 mol/kg0.9840.983−0.001
CALCULATION

How to calculate the result

Formula
Difference = reading − expected aw (at the measurement temperature)
Worked example
  1. 1

    NaCl 6.0 mol/kg standard: 0.762 − 0.760 = +0.002 aw

  2. 2

    All three differences are within ±0.003 aw

ACCEPTANCE CRITERIA

Pass if

  • Dew point instruments: typically within ±0.003 aw of the standard
  • Capacitance instruments: typically within ±0.01–0.015 aw, or the manufacturer's stated accuracy
  • Measurement temperature within the set tolerance (for example 25.0 ± 0.2 °C)
IF IT FAILS

What to do

  • Clean the chamber and sensor, then repeat with a fresh standard.
  • If still out, adjust the calibration offset as the manual allows and verify again with a second standard.
  • If adjustment fails, tag the meter out of use and send it for service.
  • Review sample results since the last passing verification.
In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how