Sulphur analyser · Calibration procedure

Calibration with sulphur standards

Builds a calibration line from standards of known sulphur content in a matching matrix, checks its linearity, and confirms it with an independent check standard.

Photo of a Sulphur analyser
Photo: Elemental analysis, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Both XRF and UVF analysers are comparative: they convert a signal to sulphur content using a calibration made from standards. A correct, matrix-matched calibration across the working range is needed so results near the 10 mg/kg specification limit (or higher limits for heavier fuels) are reliable.

Frequency

Full calibration typically when the instrument is installed, after major maintenance (new UVF combustion tube, XRF tube/detector service), when the check standard fails, or at the interval in your SOP; check standards daily.

Learn how the Sulphur analyser works
YOU WILL NEED

Standards & tools

  • Certified sulphur standards covering the range, e.g. di-n-butyl sulphide or dibenzothiophene in isooctane, toluene or mineral oil, matched to the sample type
  • Sulphur-free solvent or white oil as blank
  • Independent check standard (different lot or supplier) near the specification limit
BEFORE YOU START

Prerequisites

  • Instrument warmed up and stable; UVF gas flows and furnace temperature correct
  • Balance calibrated if standards are prepared gravimetrically
  • Syringe or autosampler checked for volume accuracy (UVF)
PROCEDURE

Step by step

  1. 1Choose a calibration range that brackets the expected samples (for example 0–50 mg/kg for ULSD by UVF).
  2. 2Measure the blank and each standard in replicate (for example three injections each for UVF).
  3. 3Fit the calibration line (signal against concentration) by least squares.
  4. 4Check the correlation coefficient and residuals of each point.
  5. 5Measure the independent check standard and calculate its recovery.
  6. 6Save the calibration and record standard lot numbers.
RECORD

Example observation table

Standard (mg/kg S)Mean response (units)
0 (blank)0.012
50.538
101.071
252.648
505.301
Check std 10.0 mg/kg1.112
CALCULATION

How to calculate the result

Formula
Response = slope × concentration + intercept; Found = (response − intercept) ÷ slope; Recovery (%) = found ÷ certified × 100
Worked example
  1. 1

    Least-squares fit of the five points: slope = 0.10576 units per mg/kg, intercept = 0.0103, r = 0.99999.

  2. 2
    Check standard: found = (1.112 − 0.0103) ÷ 0.10576 = 10.42 mg/kg.
  3. 3
    Recovery = 10.42 ÷ 10.0 × 100 = 104.2%.
  4. 4

    With an illustrative SOP limit of 90–110% at this level, the calibration is accepted.

ACCEPTANCE CRITERIA

Pass if

  • Correlation coefficient typically ≥ 0.999 (or as in your SOP), with no point showing a large residual.
  • Independent check standard within the SOP limits (for example ±10% near 10 mg/kg) or within the method's expected precision.
  • Blank response low and stable.
IF IT FAILS

What to do

  • Check standards for evaporation or contamination and prepare fresh ones.
  • UVF: check combustion tube for soot, gas flows, membrane dryer and syringe; XRF: check film, helium and tube stability.
  • Repeat the calibration; do not analyse samples until the check standard passes.