Slump cone · Calibration procedure

Dimension check

Measures the top and base internal diameters, height and wall thickness of the slump cone and checks the tamping rod, against the standard dimensions of 100 mm, 200 mm and 300 mm.

Photo of a Slump cone
Photo: Habib M’henni, CC BY 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

A dented or worn cone holds a different volume and shape of concrete, which changes the slump. Regular checks keep site and lab slump results comparable.

Frequency

Typically every 6 to 12 months for lab cones and more often for site cones (e.g. quarterly), and after any visible damage, as defined in your SOP.

Learn how the Slump cone works
YOU WILL NEED

Standards & tools

  • Calibrated steel rule or tape (1 mm)
  • Calibrated vernier calliper or internal calliper
  • Calibrated thickness gauge or micrometer
BEFORE YOU START

Prerequisites

  • Cone clean and free of hardened concrete
  • Placed on a flat surface
PROCEDURE

Step by step

  1. 1Measure the internal top diameter in two directions at right angles.
  2. 2Measure the internal base diameter in two directions at right angles.
  3. 3Measure the height at two or more positions around the cone.
  4. 4Measure the wall thickness at several points.
  5. 5Check the tamping rod diameter and length and that its end is rounded.
  6. 6Inspect for dents, seams, rust and loose foot pieces or handles.
RECORD

Example observation table

DimensionNominal (mm)Reading 1 (mm)Reading 2 (mm)Mean (mm)Result
Top internal diameter100101100100.5Pass
Base internal diameter200199201200.0Pass
Height300299300299.5Pass
Wall thickness≥ 1.5 (ASTM)1.61.61.6Pass
Tamping rod diameter1616.016.016.0Pass
CALCULATION

How to calculate the result

Formula
Mean = (Reading 1 + Reading 2) ÷ 2; Deviation = Mean − Nominal
Worked example
  1. 1
    Top diameter: (101 + 100) ÷ 2 = 100.5 mm; deviation = +0.5 mm
  2. 2
    Height: (299 + 300) ÷ 2 = 299.5 mm; deviation = −0.5 mm
  3. 3

    Both are well within a typical tolerance of ±3 mm (ASTM C143); check your IS 7320 / SOP tolerance.

ACCEPTANCE CRITERIA

Pass if

  • Top diameter 100 mm, base diameter 200 mm and height 300 mm, each within the tolerance of IS 7320 / ASTM C143 or your SOP (ASTM typically ±3 mm).
  • Wall thickness not less than the standard minimum; no dents that change the shape.
  • Tamping rod about 16 mm diameter and 600 mm long with a rounded end.
IF IT FAILS

What to do

  • Remove the cone from use and replace it.
  • Replace a bent or short tamping rod.
  • Note the issue against any slump results taken with it since the last check.