LimsPulse
Fridge / freezer · Calibration procedure

Temperature mapping

Logs temperature at several positions inside the refrigerator or freezer to find warm and cold spots and prove that the storage area stays within its range.

Photo of a Fridge / freezer
Photo: Magnus Manske, CC BY-SA 3.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Temperature near the door, the cooling plate and the top shelf can differ from the display reading. Mapping proves which positions are safe for storing material and where to place the monitoring probe.

Frequency

At qualification, then typically annually, and after relocation, major repair or a change in use, as defined in your SOP.

Learn how the Fridge / freezer works
YOU WILL NEED

Standards & tools

  • Calibrated temperature data loggers traceable to national standards (commonly 9 for a standard unit, more for large units and cold rooms)
BEFORE YOU START

Prerequisites

  • Display probe calibrated
  • Mapping protocol approved (positions, duration, interval, door-opening test)
  • Unit stable for at least 24 hours
PROCEDURE

Step by step

  1. 1Place loggers at the corners and centre of the top, middle and bottom shelves, near the door and next to the display probe.
  2. 2Log every 1–5 minutes for the protocol period, commonly 24–72 hours, empty and then loaded.
  3. 3Include a door-opening test and a defrost cycle (freezers) within the logging period.
  4. 4Optionally switch off power for a short period to measure hold-over time, if the protocol requires.
  5. 5Download data; report minimum, maximum and mean at each logger and the recovery time.
RECORD

Example observation table

LoggerPositionMin (°C)Max (°C)Mean (°C)
T1Top front3.96.84.8
T2Top rear4.16.25.1
T3Middle rear (near cooling plate)2.95.44.2
T4Middle front4.66.75.6
T5Middle centre3.85.64.5
T6Bottom rear4.25.95.0
T7Door shelf4.97.66.1
T8Bottom front3.75.84.6
T9Next to display probe4.36.35.3
CALCULATION

How to calculate the result

Overall min = lowest logger minimum; Overall max = highest logger maximum; Mean = average of logger means
  1. Range 2–8 °C
  2. Overall minimum = 2.9 °C (T3), above 2 °C
  3. Overall maximum = 7.6 °C (T7, door shelf), below 8 °C but close to the limit
  4. Mean of logger means = (4.8 + 5.1 + 4.2 + 5.6 + 4.5 + 5.0 + 6.1 + 4.6 + 5.3) ÷ 9 = 45.2 ÷ 9 = 5.02 °C
  5. The unit passes; the door shelf is the warmest position and should not be used for critical material
ACCEPTANCE CRITERIA

Pass if

  • All loggers within the storage range for the whole study: typically 2–8 °C for a refrigerator and as defined (e.g. −25 °C to −10 °C, or the product label range) for a freezer.
  • Recovery after the door-opening test within the time in the protocol.
IF IT FAILS

What to do

  • Restrict storage to positions that passed, or remove the unit from use.
  • Service the unit (door seal, fan, controller) and repeat mapping.
  • Assess the impact on material stored since the last passing mapping.
In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how