pH meter · Calibration procedure

Temperature compensation

Checks that the temperature probe reads correctly and that the meter's automatic temperature compensation gives the right pH at a temperature other than 25 °C.

Photo of a pH meter
Photo: Datamax, Public domain (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

The electrode slope changes with temperature (59.16 mV/pH at 25 °C, about 60.15 mV/pH at 30 °C). If the temperature reading is wrong, the meter uses the wrong slope and reports the wrong pH, especially far from pH 7.

Frequency

Typically every 6 months or yearly, and when a new electrode or temperature probe is fitted, as defined in your calibration SOP.

Learn how the pH meter works
YOU WILL NEED

Standards & tools

  • Calibrated reference thermometer (traceable certificate)
  • Certified pH buffer with a temperature table on its certificate, e.g. pH 10.01 buffer
  • Thermostatic water bath
BEFORE YOU START

Prerequisites

  • Meter calibrated with buffers at 25 °C and ATC switched on
  • Water bath set to two temperatures, e.g. about 20 °C and 30 °C
PROCEDURE

Step by step

  1. 1Place the ATC probe and the reference thermometer side by side in the water bath at about 20 °C. Record both readings when stable.
  2. 2Repeat at about 30 °C (or another temperature used by your methods).
  3. 3Place a beaker of pH 10.01 buffer in the bath at 20 °C, measure its pH with the electrode, and compare with the certificate value at 20 °C.
  4. 4Repeat with the buffer at 30 °C.
RECORD

Example observation table

Bath (°C, reference)ATC probe (°C)Difference (°C)Buffer certificate pHObserved pHDifference (pH)
20.120.3+0.210.0610.07+0.01
30.029.8−0.29.979.96−0.01
CALCULATION

How to calculate the result

Theoretical slope (mV/pH) = 0.19841 × (t + 273.15), where t is in °C
  1. At 25 °C: 0.19841 × 298.15 = 59.16 mV/pH
  2. At 30 °C: 0.19841 × 303.15 = 60.15 mV/pH
  3. Temperature difference at 30 °C = 29.8 − 30.0 = −0.2 °C
ACCEPTANCE CRITERIA

Pass if

  • ATC probe within ±0.5 °C of the reference thermometer (or the limit in your SOP).
  • Buffer reading within ±0.05 pH of the certificate value at that temperature.
IF IT FAILS

What to do

  • Check the probe connection and that ATC is switched on; repeat.
  • Apply a temperature offset only if the meter software allows it and your SOP permits.
  • Otherwise replace the temperature probe or electrode and repeat the check.