LimsPulse
Osmometer · Calibration procedure

Calibration with osmolality standards

Calibrates the osmometer with certified sodium chloride osmolality standards that bracket the working range, then checks it with a reference solution.

Photo of a Osmometer
Photo: Zemxer, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

The osmometer turns a freezing point into mOsm/kg using its calibration line. Calibrating with certified standards makes sure the conversion is correct over the range used for products.

Frequency

Typically every 6 or 12 months, after repair or probe replacement, or when the reference check fails. A reference check is usually run daily. As defined in your SOP.

Learn how the Osmometer works
YOU WILL NEED

Standards & tools

  • Certified osmolality standards (usually sodium chloride in water), e.g. 100 mOsm/kg and 500 mOsm/kg (or as the manufacturer specifies), plus a 290 mOsm/kg reference solution
  • Calibrated precision pipette
BEFORE YOU START

Prerequisites

  • Instrument at operating temperature
  • Probe clean and dry
  • Fresh ampoules of standards within expiry; opened ampoules used straight away
PROCEDURE

Step by step

  1. 1Enter the calibration mode.
  2. 2Measure the low standard (e.g. 100 mOsm/kg) as the instrument directs; repeat if the software requires.
  3. 3Measure the high standard (e.g. 500 mOsm/kg).
  4. 4Accept the calibration when the software confirms it.
  5. 5Measure the 290 mOsm/kg reference solution in triplicate as a check.
  6. 6Measure each standard again as a sample to confirm accuracy, if your SOP requires.
RECORD

Example observation table

Standard (mOsm/kg)Reading 1Reading 2Reading 3MeanDifferenceResult
100101100101100.7+0.7Pass
290 (reference)289290289289.3−0.7Pass
500498499500499.0−1.0Pass
CALCULATION

How to calculate the result

Difference = Mean reading − Certified value; expected freezing point depression ΔTf = 1.86 × osmolality (osmol/kg)
  1. 290 mOsm/kg reference: mean = (289 + 290 + 289) ÷ 3 = 289.3 mOsm/kg; difference = −0.7
  2. Expected ΔTf for 290 mOsm/kg = 1.86 × 0.290 = 0.539 °C
  3. 500 mOsm/kg standard: mean = 499.0; difference = −1.0
ACCEPTANCE CRITERIA

Pass if

  • Each standard and the reference solution within the limit in your SOP, often based on the manufacturer's specification (for example within about ±2 mOsm/kg near 290 mOsm/kg, or ±0.5% to 1% at higher values).
  • Repeatability within the manufacturer's specification.
IF IT FAILS

What to do

  • Clean the probe, use fresh ampoules of standards and repeat the calibration.
  • Check the pipette and the sample volume.
  • If it still fails, call service and assess results since the last passing check.
In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how