Mass accuracy calibration

Calibrates the m/z scale of the mass spectrometer with a reference solution, then verifies accuracy on known compounds.

Photo of a LC-MS
Photo: Sarka Na kopci, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Identification by MS depends on measuring the right mass. A drifting mass scale can mis-assign ions, lower sensitivity in MRM methods or give a wrong elemental formula on high-resolution instruments.

Frequency

High-resolution instruments are typically calibrated or checked daily or before each batch; quadrupole instruments typically every few months and after maintenance or venting, as defined in your SOP and the manufacturer's guide.

Learn how the LC-MS works
YOU WILL NEED

Standards & tools

  • Manufacturer's calibration solution covering the working mass range (e.g. sodium formate or caesium iodide clusters, or a proprietary tuning mix)
  • Check compounds with known exact masses, e.g. caffeine, leucine enkephalin, reserpine
  • LC-MS grade solvents
BEFORE YOU START

Prerequisites

  • Vacuum stable and the instrument at operating temperature for the time the manufacturer specifies
  • Ion source clean; spray stable when infusing calibrant
  • Room temperature within the instrument's specified range
PROCEDURE

Step by step

  1. 1Infuse the calibration solution in the polarity (positive and/or negative) to be calibrated.
  2. 2Run the manufacturer's calibration routine; it fits the measured peaks to their known m/z values.
  3. 3Review the calibration report: number of ions used, residual error of each and overall fit.
  4. 4Inject or infuse the check compounds and record their measured m/z.
  5. 5Calculate mass error for each check ion and compare with the limit.
RECORD

Example observation table

Check ionTheoretical m/zMeasured m/zError (ppm)
Caffeine [M+H]⁺195.0877195.0874−1.5
Leucine enkephalin [M+H]⁺556.2766556.2771+0.9
Reserpine [M+H]⁺609.2807609.2815+1.3
CALCULATION

How to calculate the result

Mass error (ppm) = (Measured m/z − Theoretical m/z) ÷ Theoretical m/z × 10⁶; for unit-resolution instruments, error (Da) = Measured − Theoretical
  1. Reserpine [M+H]⁺: theoretical 609.2807, measured 609.2815
  2. Difference = 609.2815 − 609.2807 = 0.0008
  3. Error = 0.0008 ÷ 609.2807 × 10⁶ = +1.3 ppm
ACCEPTANCE CRITERIA

Pass if

  • High-resolution MS (Q-TOF, Orbitrap): typically within ±5 ppm with external calibration (often ±2–3 ppm, or better with a lock mass), as per manufacturer and SOP.
  • Unit-resolution quadrupole MS: typically within ±0.1–0.2 Da of the theoretical value, as per manufacturer and SOP.
IF IT FAILS

What to do

  • Clean the ion source and check spray stability, then recalibrate.
  • Check room temperature and allow the analyser to stabilise.
  • Do not use the instrument for identification or quantitation until it passes.
  • Contact service if the calibration repeatedly fails.