Friability tester · Calibration procedure

Rotation speed & count

Checks the drum turns at 25 ± 1 rpm and stops after the set number of revolutions.

Illustration of Friability tester
PURPOSE

Why this is checked

Friability is a fixed amount of tumbling. If the speed or the count is wrong, tablets receive more or less stress than intended.

Frequency

Typically every 6 months, as defined in your calibration SOP.

Learn how the Friability tester works
YOU WILL NEED

Standards & tools

  • Calibrated stopwatch
  • Optional calibrated tachometer
BEFORE YOU START

Prerequisites

  • Drum clean and fitted
  • Counter set to 100 revolutions
PROCEDURE

Step by step

  1. 1Mark a reference point on the drum.
  2. 2Start the drum with the counter set to 100 and the stopwatch running.
  3. 3Count the revolutions by watching the mark and stop the stopwatch when the drum stops.
  4. 4Calculate rpm = revolutions ÷ time in minutes.
  5. 5Repeat three times (and for each drum on dual-drum models).
RECORD

Example observation table

TrialSet revolutionsCounted revolutionsTime (s)Speed (rpm)
1100100240.524.95
2100100239.625.04
3100100240.224.98
CALCULATION

How to calculate the result

Speed (rpm) = Revolutions ÷ (Time in seconds ÷ 60)
  1. Trial 1: 100 revolutions in 240.5 s
  2. Speed = 100 ÷ (240.5 ÷ 60) = 100 ÷ 4.008 = 24.95 rpm (within 25 ± 1 rpm)
ACCEPTANCE CRITERIA

Pass if

  • Speed 25 ± 1 rpm.
  • Counted revolutions equal to the set number (e.g. 100).
IF IT FAILS

What to do

  • Label out of calibration
  • Service the motor or counter
  • Recalibrate before use