DSC · Calibration procedure

Temperature and enthalpy calibration (indium)

Melts a certified indium standard and checks the onset temperature and heat of fusion against the certified values (onset 156.6 °C, ΔH about 28.45 J/g).

Photo of a DSC
Photo: Nick Birse, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

DSC temperature and peak area depend on the sensor calibration. Indium melts sharply at a known temperature with a known enthalpy, so it checks both the temperature axis and the heat-flow (cell constant) in one run.

Frequency

Typically a check each week or each month of use and a full calibration every 6–12 months, or after repair, as defined in your SOP.

Learn how the DSC works
YOU WILL NEED

Standards & tools

  • Certified indium reference material (onset about 156.6 °C, enthalpy of fusion about 28.45 J/g; use the certificate values)
  • Optional second standard for wider ranges, for example zinc (melting about 419.5 °C) or tin
  • Calibrated microbalance
BEFORE YOU START

Prerequisites

  • Purge gas and flow as in routine methods (for example nitrogen 50 mL/min)
  • Same pan type and heating rate as routine use (for example 10 °C/min)
  • Clean cell and a flat baseline
PROCEDURE

Step by step

  1. 1Weigh about 5–10 mg of indium accurately into an aluminium pan and crimp the lid.
  2. 2Place it in the cell with an empty reference pan.
  3. 3Heat from about 120 °C to 180 °C at the routine heating rate.
  4. 4Evaluate the extrapolated onset temperature and the peak area (J/g) with a linear baseline.
  5. 5If adjustment is needed, enter the measured values in the calibration software, then rerun to confirm.
RECORD

Example observation table

ParameterCertified valueObservedDifference
Onset temperature (°C)156.6156.8+0.2 °C
Enthalpy of fusion (J/g)28.4528.21−0.84%
CALCULATION

How to calculate the result

ΔT = Observed onset − Certified onset; % error ΔH = (Observed ΔH − Certified ΔH) ÷ Certified ΔH × 100; Cell constant K = Certified ΔH ÷ Observed ΔH
  1. ΔT = 156.8 − 156.6 = +0.2 °C
  2. % error ΔH = (28.21 − 28.45) ÷ 28.45 × 100 = −0.84%
  3. K = 28.45 ÷ 28.21 = 1.0085
ACCEPTANCE CRITERIA

Pass if

  • Onset within about ±0.5 °C of the certified value (typical; check your SOP).
  • Enthalpy within about ±2% of the certified value (some SOPs use ±1%).
IF IT FAILS

What to do

  • Check the pan sits flat, the purge gas is correct and the indium is clean.
  • Recalibrate the temperature and cell constant using the manufacturer's procedure and rerun indium.
  • Assess the impact on results generated since the last passing calibration.
  • Repeat the full calibration before returning it to use.