LimsPulse
DO meter · Calibration procedure

Air and zero calibration

Sets the DO sensor's 100% point in water-saturated air and its 0% point in oxygen-free solution, then checks the reading in air-saturated water against the solubility table.

Photo of a DO meter
Photo: Nuno Nogueira (Nmnogueira), CC BY-SA 2.5 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

DO sensors drift as membranes, electrolyte or sensor caps age. A two-point calibration fixes the slope and zero; the check in air-saturated water confirms the meter reads true mg/L at the actual temperature and pressure.

Frequency

Air calibration typically daily or before use; zero calibration weekly or monthly, or as defined in your SOP.

Learn how the DO meter works
YOU WILL NEED

Standards & tools

  • Water-saturated air (calibration sleeve or bottle with a little water)
  • Zero-oxygen solution, for example freshly prepared sodium sulphite solution (about 2% w/v in water, sometimes with a trace of cobalt salt as catalyst)
  • Calibrated barometer (or the meter's built-in barometer) and reference thermometer
  • Oxygen solubility table (for example from ISO 5814)
BEFORE YOU START

Prerequisites

  • Membrane or sensor cap in good condition; electrolyte filled (electrochemical sensors)
  • Probe polarised for the time stated by the supplier
  • Probe and calibration vessel at a stable temperature
PROCEDURE

Step by step

  1. 1Place the probe in the calibration sleeve with water-saturated air, out of direct sun and draughts.
  2. 2Enter or confirm the barometric pressure and set salinity to zero for fresh water.
  3. 3When the reading is stable, accept the 100% (air) calibration.
  4. 4Rinse the probe and place it in the zero-oxygen solution; when stable, accept the zero calibration.
  5. 5Rinse the probe thoroughly with purified water.
  6. 6Check: measure air-saturated water (water stirred or aerated until saturated) and compare with the theoretical value from the solubility table corrected for pressure.
RECORD

Example observation table

CheckConditionExpectedReadingDifference
ZeroSodium sulphite solution0.00 mg/L0.03 mg/L+0.03 mg/L
Air-saturated water25.0 °C, 985 hPa8.03 mg/L8.01 mg/L−0.02 mg/L
CALCULATION

How to calculate the result

Expected DO (mg/L) = DO at saturation from the table (at 1013.25 hPa) × (Barometric pressure ÷ 1013.25 hPa)
  1. Saturation DO in fresh water at 25 °C and 1013.25 hPa ≈ 8.26 mg/L (from the solubility table)
  2. Barometric pressure = 985 hPa
  3. Expected = 8.26 × 985 ÷ 1013.25 = 8.03 mg/L
  4. Reading 8.01 mg/L → difference −0.02 mg/L
ACCEPTANCE CRITERIA

Pass if

  • Zero reading typically not more than about 0.1 mg/L.
  • Air-saturated water reading typically within ±0.2 mg/L (or about ±2% saturation) of the expected value, or the limits in your SOP.
  • Sensor slope or calibration value within the range given by the supplier.
IF IT FAILS

What to do

  • Check pressure and salinity settings and that the probe is fully polarised.
  • Replace the membrane and electrolyte, or the sensor cap, and recalibrate.
  • Prepare fresh zero solution.
  • Label the meter out of calibration if it still fails.
In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how