LimsPulse
Density meter · Calibration procedure

Air and water check

Measures dry air and pure water in the U-tube at the measuring temperature and compares each reading with its known density.

Photo of a Density meter
Photo: MTan42, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

The density meter calculates density from two constants found with air and water. Checking both shows whether the tube is clean, dry and still correctly adjusted.

Frequency

Typically daily or before use, and after cleaning or maintenance, as defined in your SOP. A full calibration with certified standards is usually done every 6 or 12 months.

Learn how the Density meter works
YOU WILL NEED

Standards & tools

  • Dry air (density calculated by the instrument from the barometric pressure and temperature, about 0.00120 g/cm³ at 20 °C and 1013 hPa)
  • Freshly boiled and cooled purified water (0.99820 g/cm³ at 20 °C)
BEFORE YOU START

Prerequisites

  • Tube cleaned and dried
  • Temperature set to 20.00 °C and stable
  • Barometric pressure entered or measured by the instrument
PROCEDURE

Step by step

  1. 1With the tube dry, start the air check. Record the reading and the expected air density.
  2. 2Fill the tube slowly with degassed purified water, checking for bubbles.
  3. 3Start the water check and record the reading.
  4. 4Compare each reading with the expected value.
  5. 5Rinse and dry the tube.
RECORD

Example observation table

StandardExpected (g/cm³ at 20 °C)Observed (g/cm³)Difference (g/cm³)Result
Dry air (1013 hPa)0.001200.00121+0.00001Pass
Purified water0.998200.99822+0.00002Pass
CALCULATION

How to calculate the result

Difference = Observed − Expected
  1. Air density at 20 °C and 101.325 kPa = p ÷ (R × T) = 101325 ÷ (287.05 × 293.15) = 1.204 kg/m³ = 0.00120 g/cm³
  2. Air: 0.00121 − 0.00120 = +0.00001 g/cm³
  3. Water: 0.99822 − 0.99820 = +0.00002 g/cm³
ACCEPTANCE CRITERIA

Pass if

  • Water within the limit in your SOP, commonly ±0.0001 g/cm³ for a 4-decimal instrument (tighter for higher-resolution models).
  • Air within the limit set by the manufacturer or your SOP.
IF IT FAILS

What to do

  • Clean the tube thoroughly, dry it and repeat with fresh degassed water.
  • If it still fails, perform an air and water adjustment as the manufacturer describes and check again with a certified density standard.
  • If it fails after adjustment, call service and assess results since the last pass.
In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how