LimsPulse
Centrifuge · Calibration procedure

Timer verification

Checks the centrifuge run timer against a calibrated stopwatch.

Photo of a Centrifuge
Photo: rebcenter-moscow.ru, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Methods specify a centrifugation time. A timer that runs short gives incomplete separation; one that runs long wastes time and may warm samples.

Frequency

Typically together with speed verification, every 6 or 12 months, as defined in your calibration SOP.

Learn how the Centrifuge works
YOU WILL NEED

Standards & tools

  • Calibrated stopwatch traceable to national standards
BEFORE YOU START

Prerequisites

  • Rotor balanced and lid closed
  • Know whether the timer counts from start or from reaching set speed (check the manual)
PROCEDURE

Step by step

  1. 1Set the timer to a short and a long time used in methods, e.g. 10 and 30 minutes.
  2. 2Start the stopwatch at the same moment as the run (or when set speed is reached, depending on the timer mode).
  3. 3Stop the stopwatch when the timer ends (deceleration starts).
  4. 4Record the measured time and calculate the % error.
RECORD

Example observation table

Set timeSet (s)Stopwatch (s)% error
10 min600603+0.50
30 min18001802+0.11
CALCULATION

How to calculate the result

% error = (Measured time − Set time) ÷ Set time × 100
  1. Set 10 min = 600 s; stopwatch reads 603 s
  2. % error = (603 − 600) ÷ 600 × 100 = +0.50%
ACCEPTANCE CRITERIA

Pass if

  • Timer typically within ±2% of the set time (or a fixed limit such as ±10 s), as defined in your SOP.
IF IT FAILS

What to do

  • Label the centrifuge and arrange service of the controller.
  • Use a calibrated external timer temporarily only if your SOP allows.
MORE

Other Centrifuge calibrations

In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how