Cell counter · Calibration procedure
Counting accuracy with reference beads
Counts certified polystyrene beads of known concentration and size several times, and checks the mean count, precision and size against the certificate.

PURPOSE
Why this is checked
Frequency
Typically monthly or quarterly, after service and before validation work, as defined in your SOP.
Learn how the Cell counter worksYOU WILL NEED
Standards & tools
- Certified concentration and size bead standard (for example about 1 × 10⁶ beads/mL, nominal size in the cell range)
- Bead certificate stating mean concentration with its uncertainty and mean size
BEFORE YOU START
Prerequisites
- Counter clean and self-test passed
- Beads in date, at room temperature and fully resuspended (vortex as instructed)
- Bead protocol selected with dilution factor 1 (no stain) unless the supplier says otherwise
PROCEDURE
Step by step
- 1Vortex the bead vial for the time stated on the certificate.
- 2Load a slide with beads and count. Repeat to give at least 5 independent counts from fresh loads.
- 3Record concentration and mean size for each count.
- 4Calculate mean, standard deviation, % CV and recovery against the certified value.
- 5Compare the mean size with the certified size.
RECORD
Example observation table
| Replicate | Count (beads/mL) |
|---|---|
| 1 | 1.02 × 10⁶ |
| 2 | 0.98 × 10⁶ |
| 3 | 1.05 × 10⁶ |
| 4 | 1.01 × 10⁶ |
| 5 | 0.99 × 10⁶ |
| Mean / SD / CV | 1.010 × 10⁶ / 0.027 × 10⁶ / 2.7% |
CALCULATION
How to calculate the result
Formula
Recovery (%) = mean count ÷ certified count × 100; CV (%) = SD ÷ mean × 100
Worked example
- 1Mean = (1.02 + 0.98 + 1.05 + 1.01 + 0.99) ÷ 5 = 1.010 × 10⁶ beads/mL
- 2SD = 0.027 × 10⁶; CV = 0.027 ÷ 1.010 × 100 = 2.7%
- 3Certified value 1.00 × 10⁶ beads/mL → recovery = 1.010 ÷ 1.00 × 100 = 101.0%
ACCEPTANCE CRITERIA
Pass if
- Mean count within the certificate range, typically about ±10% of the certified value
- CV typically ≤ 5–10%, as set in your SOP
- Mean size within the tolerance stated on the certificate
IF IT FAILS
What to do
- Check bead resuspension and slide loading; repeat with a new vial if needed.
- If still failing, label the counter out of use for reportable counts and call service.
- Assess counts made since the last passing check.
In LimsPulse · Calibration Management
LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.