Finished products · Injections (small-volume and large-volume parenterals)

Particulate Matter in Injections by Light Obscuration

Counts sub-visible particles of 10 µm and above and 25 µm and above in an injection using a light obscuration counter, and compares the counts per container (or per mL) with pharmacopoeial limits.

  • Light obscuration particle count test
  • Liquid particle counter

Reference: Typical approach based on USP <788> method 1 and Ph. Eur. 2.9.19. Always follow the current chapter and your approved STP.

PRINCIPLE

Why this method works

The injection is drawn through the narrow sensor of a light obscuration counter. Each particle blocks part of a light beam, giving a pulse whose height depends on its size. Pulses are counted into size classes calibrated with certified spheres. Several portions are measured, the first is discarded and the rest are averaged. The average count per portion is scaled to particles per container (small-volume injections) or per mL (large-volume injections).

Instruments used

MATERIALS

Reagents & standards

ReagentGrade / specification
Particle-free waterPurified water filtered through a membrane of 1.2 µm or finer
GlasswareCleaned and rinsed with particle-free water
SAMPLE PREPARATION

Preparing the sample

  1. 01

    Blank

    Run particle-free water to show the environment and glassware are clean.

  2. 02

    Mix

    Invert each container gently about 20 times.

  3. 03

    Pool

    For small units, pool enough containers to give at least 25 mL.

  4. 04

    Degas

    Let stand or apply gentle vacuum to remove bubbles.

  5. 05

    Measure

    Run four portions of not less than 5 mL each.

  6. 06

    Calculate

    Discard the first; average the rest and scale per container or per mL.

Procedure

  1. 1Carry out the test under a laminar air flow cabinet wearing powder-free gloves.
  2. 2Run a blank of particle-free water (e.g. 5 portions of 5 mL); it should meet the environment limit (commonly not more than 25 particles ≥ 10 µm in 25 mL).
  3. 3Wipe the outside of each container. Mix by inverting gently about 20 times; do not shake.
  4. 4For containers of 25 mL or more, test units individually. For smaller units, open and pool the contents of enough units (commonly 10 or more) into a clean vessel to give at least 25 mL.
  5. 5Let the sample stand, or apply gentle vacuum, until no air bubbles are visible.
  6. 6Measure four portions of not less than 5 mL each and record counts at ≥ 10 µm and ≥ 25 µm.
  7. 7Discard the first portion and calculate the mean of the remaining portions.

Standard preparation

  1. 1No standard is run with each test; sizing calibration, sensor resolution, count accuracy and volume accuracy are verified at the calibration interval.

Notes for analysts

  • Most high results come from contamination during handling; repeat the blank before blaming the product.
  • Large-volume results are per mL: average count per portion ÷ portion volume.
  • Protein products may contain translucent particles that light obscuration undersizes; orthogonal methods (e.g. flow imaging) are used in development.
TEST CONDITIONS

Method parameters (typical)

InstrumentLight obscuration particle counter calibrated with certified polystyrene spheres
Size channels≥ 10 µm and ≥ 25 µm
PortionsFour portions of not less than 5 mL; first discarded
EnvironmentLaminar air flow; particle-free water blank
Limits (small-volume, ≤ 100 mL)Typically not more than 6000 particles ≥ 10 µm and 600 particles ≥ 25 µm per container (see current chapter)
Limits (large-volume, > 100 mL)Typically not more than 25 particles/mL ≥ 10 µm and 3 particles/mL ≥ 25 µm (see current chapter)
BEFORE YOU REPORT

System suitability & sequence

ParameterLimit
Blank (environment)Typically not more than 25 particles ≥ 10 µm in 25 mL of particle-free water
Instrument calibrationSizing, resolution, count accuracy and volume accuracy within due date

Typical sequence

  1. 1Particle-free water blank
  2. 2Sample pool / unit: portion 1 (discard)
  3. 3Portions 2–4 (average)
  4. 4Next sample
CALCULATION

Calculation with a worked example

Particles per container = (A × VT) ÷ (VA × n)

Where

A
Mean particle count of the portions analysed (portions 2–4)
VT
Total volume of the pooled sample (mL)
VA
Volume of each portion analysed (mL)
n
Number of containers pooled

Given

Product
2 mL ampoules; 15 ampoules pooled
VT
30 mL
VA
5 mL
n
15
Counts ≥ 10 µm (portions 2–4)
183, 171, 177
Counts ≥ 25 µm (portions 2–4)
6, 4, 5

Working

  1. ≥ 10 µm: A = (183 + 171 + 177) ÷ 3 = 177 per 5 mL
  2. Per container = (177 × 30) ÷ (5 × 15) = 5310 ÷ 75 = 70.8
  3. ≥ 25 µm: A = (6 + 4 + 5) ÷ 3 = 5 per 5 mL
  4. Per container = (5 × 30) ÷ (5 × 15) = 150 ÷ 75 = 2.0

≈ 71 particles ≥ 10 µm and 2 particles ≥ 25 µm per container, well within the typical small-volume limits of 6000 and 600 per container: complies.

ACCEPTANCE CRITERIA

What result passes

  • Small-volume injections (≤ 100 mL): typically not more than 6000 particles ≥ 10 µm and not more than 600 particles ≥ 25 µm per container.
  • Large-volume injections (> 100 mL): typically not more than 25 particles per mL ≥ 10 µm and not more than 3 particles per mL ≥ 25 µm.
  • If the light obscuration test fails, USP <788> allows testing by the microscopic particle count method (method 2). Always check the current chapter.
METHOD VALIDATION

Validation parameters (ICH Q2(R2))

ParameterTypical requirement
SuitabilityProduct does not interfere (colour, viscosity, bubbles, precipitation); otherwise dilute or use the microscopic method
PrecisionRepeat counts on the same pool agree
RobustnessMixing, degassing and pooling steps shown not to add or remove particles