LimsPulse
Turbidity meter · Calibration procedure

Formazin standard calibration

Calibrates the turbidity meter with a series of formazin standards covering the working range, then verifies the calibration with an independent standard.

Illustration of Turbidity meter
PURPOSE

Why this is checked

Formazin is the primary reference for turbidity. Calibrating with it fixes the relationship between the detector signal and NTU, so sample results are correct and comparable.

Frequency

Typically every 3 to 12 months as the manufacturer and your SOP require, after lamp replacement or repair, and whenever daily verification fails.

Learn how the Turbidity meter works
YOU WILL NEED

Standards & tools

  • Formazin stock suspension, 4000 NTU (prepared from hydrazine sulphate and hexamethylenetetramine, or purchased stabilised formazin)
  • Dilution water: purified water filtered through a 0.2 µm membrane
  • Calibrated volumetric pipettes and flasks
  • Independent verification standard (e.g. stabilised formazin or secondary gel standard)
BEFORE YOU START

Prerequisites

  • Instrument warmed up
  • Cuvettes clean, unscratched and indexed
  • Standards freshly diluted on the day of use (dilute formazin is not stable)
PROCEDURE

Step by step

  1. 1Prepare the calibration standards by diluting the 4000 NTU stock, for example 20, 100 and 800 NTU (or the levels the manufacturer specifies).
  2. 2Measure the dilution water as the blank (it should read very low, typically below about 0.1 NTU).
  3. 3Mix each standard gently by inversion, fill a cuvette, wipe it and measure it in the calibration mode, from lowest to highest.
  4. 4Accept the calibration when the software confirms it.
  5. 5Measure the independent verification standard and record the result.
RECORD

Example observation table

StandardPreparationNominal (NTU)Verification reading (NTU)% differenceResult
Dilution water0.2 µm-filtered water< 0.10.05—Pass
Std 15 mL stock → 1000 mL2019.6−2.0%Pass
Std 225 mL stock → 1000 mL100101.2+1.2%Pass
Std 3200 mL stock → 1000 mL800792−1.0%Pass
CALCULATION

How to calculate the result

C2 = C1 × V1 ÷ V2; % difference = (Reading − Nominal) ÷ Nominal × 100
  1. 20 NTU standard: 4000 NTU × 5 mL ÷ 1000 mL = 20 NTU
  2. 800 NTU standard: 4000 × 200 ÷ 1000 = 800 NTU
  3. Verification reading 19.6 NTU: (19.6 − 20) ÷ 20 × 100 = −2.0%
ACCEPTANCE CRITERIA

Pass if

  • Each verification reading within the limit set by the manufacturer or your SOP (commonly within ±5% to ±10% of the nominal value, with a tighter absolute limit at low NTU).
  • Blank (dilution water) within the low limit in your SOP.
IF IT FAILS

What to do

  • Clean or replace cuvettes, prepare fresh dilutions and repeat.
  • Check the stock suspension's age and the dilution water.
  • If it still fails, check or replace the lamp, call service, and assess results since the last pass.