Liquid particle counter · Calibration procedure
Sizing calibration with certified spheres
Calibrates the size thresholds of the light obscuration sensor using monodisperse polystyrene spheres of certified diameter, and checks sensor resolution.
PURPOSE
Why this is checked
The counter does not measure size directly; it measures how much light each particle blocks. Certified spheres link signal height to size so the 10 µm and 25 µm channels count the right particles.
Frequency
At least annually, or as the current chapter and your SOP require, and after sensor replacement or repair.
Learn how the Liquid particle counter worksYOU WILL NEED
Standards & tools
- Certified monodisperse polystyrene spheres of several sizes covering the range (e.g. about 5, 10, 15 and 25 µm), traceable to a national standard
- Particle-free water
BEFORE YOU START
Prerequisites
- Sensor flushed and blank counts low
- Sample volume accuracy and flow rate already verified
PROCEDURE
Step by step
- 1Dilute each sphere suspension in particle-free water to a concentration well below the coincidence limit.
- 2Run each suspension and record the pulse-height (or channel) at the peak of the response.
- 3Build the calibration curve of signal height against sphere diameter in the software.
- 4Check sensor resolution with the 10 µm spheres from the width of the response distribution.
- 5Re-run the 10 µm spheres as unknowns and confirm the indicated size.
RECORD
Example observation table
| Certified sphere (µm) | Indicated mean size (µm) | % error |
|---|---|---|
| 5.0 | 5.1 | +2.0 |
| 10.0 | 10.2 | +2.0 |
| 15.0 | 14.8 | −1.3 |
| 25.0 | 25.3 | +1.2 |
CALCULATION
How to calculate the result
% sizing error = (Indicated size − Certified size) ÷ Certified size × 100
- 10.0 µm spheres read as 10.2 µm
- % error = (10.2 − 10.0) ÷ 10.0 × 100 = +2.0%
ACCEPTANCE CRITERIA
Pass if
- Indicated size of the certified spheres within the tolerance in your SOP and the current chapter.
- Sensor resolution typically not more than 10% (see USP <788> / Ph. Eur. 2.9.19 for the exact procedure).
IF IT FAILS
What to do
- Clean the sensor and repeat the calibration.
- If it still fails, label the counter 'Out of calibration' and contact service; the sensor may need replacement.
- Assess particulate results reported since the last valid calibration.